Kinematic and experimental investigation of manual resection tools for transurethral bladder tumor resection.
Аннотация
BACKGROUND
Transurethral Resection of Bladder Tumors (TURBT) is a challenging procedure partly due to resectoscope limitations. To date, manual resection performance has not been fully characterized. This work characterizes manual resection performance in the bladder while analyzing the effect of resection location on accuracy.
METHODS
Kinematic simulations are used to assess kinematic measures of resection dexterity. An experimental protocol for manual resection accuracy assessment is developed. Cross correlations between the theoretical performance measures and the observed experimental accuracy are investigated.
RESULTS
Tangential accuracy correlates relatively strongly with normal singular value and moderately with tangential kinematic conditioning index and tangential minimum singular value. Simulations also clarified difficulties in resecting close to the bladder neck.
CONCLUSIONS
Measures to evaluate accuracy and dexterity of TURBT from a kinematic viewpoint are presented to provide a currently missing quantified dexterity baseline in manual TURBT. Limitations in various bladder regions are illustrated. Copyright © 2016 John Wiley & Sons, Ltd.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.