BACKGROUND

The features of the hypertrophic cardiomyopathy (HCM) ECG make it a challenge for subcutaneous implantable cardioverter-defibrillator (S-ICD) screening. We aimed to investigate the causes of screening failure at rest and on exercise to inform optimal S-ICD ECG vector development.

METHODS AND

RESULTS

One hundred and thirty-one HCM patients (age, 50±16 years; 92 males and 39 females) with ≥1 HCM risk factor for sudden death underwent S-ICD ECG screening at rest and on exercise. Fifty patients (38%) were ineligible for S-ICD because of screening failure in every lead vector: 33 (66%) failed in the supine position, 12 (24%) failed in the standing position, and 5 (10%) failed on exercise. In patients who could exercise and passed screening at rest, 31 (44%) had 1 vector safety, 16 (23%) had 2 vector safety, and 24 (33%) had 3 vector safety. Increased R:T wave ratio in the S-ICD screening ECG (odds ratio, 4.0; confidence interval, 3.0-5.3; P<0.001) was associated with screening failure, while R/T ratio <3 in aVF (odds ratio, 0.3; confidence interval, 0.12-0.69; P=0.006) and increasing age (odds ratio, 0.97; confidence interval, 0.95-0.99; P=0.03) was associated with reduced screening failure. European Society of Cardiology risk score was higher in those failing screening (risk score 5.5% [interquartile range, 3.2-8.7] in failed versus 4.5% [interquartile range, 2.9-7.4] in passed; P=0.04).

CONCLUSIONS

HCM patients have a significant incidence of screening failure, which is determined primarily by the increased R:T ratio on the screening ECG and lead aVF. High-risk patients have an increased screening failure rate. Optimization of sensing algorithms is required to ensure that the highest risk HCM patients can benefit from S-ICD implantation.

Для цитирования: Srinivasan Neil T, Patel Kiran H, Qamar Kashif, Taylor Amy, Bacà Marco, Providência Rui, Tome-Esteban Maria, Elliott Perry M, Lambiase Pier D. Disease Severity and Exercise Testing Reduce Subcutaneous Implantable Cardioverter-Defibrillator Left Sternal ECG Screening Success in Hypertrophic Cardiomyopathy.. Circulation. Arrhythmia and electrophysiology. 2017. DOI: 10.1161/circep.117.004801.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.