Tissue Structure and Edema Fluid Events During Treatment of Lymphedema of Limbs with a Manual Pressure-Calibrated Device, Linforoll.
Аннотация
Background
Linforoll is a device composed of handpiece with roller and pressure sensor connected wireless to the computer displaying the pressure curve of the applied force. In a previous study, we proved it to regulate the applied force according to the hydromechanic conditions of the massaged tissues. Standardization of massage based on applied force was repeatable in the same patient; it decreased limb volume and provided evident increase in tissue elasticity. Methods and
Results
In this study, we measured additional parameters useful for the understanding of tissue and fluid events and approval of the device for general practice. These were skin stiffness, subcutaneous tissue stiffness independent of skin, skin water concentration, changes in skin temperature, skin capillary blood flow, subcutaneous tissue fluid pressure, volume of the moved edema fluid, and visualization of movement on indocyanine green (ICG) lymphography. Measurements were done before and during the massage. The data were obtained from a group of 20 patients with obstructive lymphedema of lower limbs during the Linforoll massage. There was a lack of significant changes in skin stiffness, skin water concentration, skin surface temperature, and capillary blood flow, but evident increase in the subcutaneous tissue elasticity (tonometry) and lymphography-shown flow of the edema fluid.
Conclusions
The skin tissue hydromechanic parameters remained normal proving lack of destructive changes under high massaging pressures. The obtained data evidently show that not the skin but the subcutis accumulated edema fluid that can successfully be moved proximally under pressures of 80-120 mmHg.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.