A Prospective Multicenter Assessment of the Accuracy and Safety of the Yuwell CT3 Real-Time Continuous Glucose Monitoring System in Patients With Diabetes Over 14 Days.
СтатьяМногоцентровое исследованиеПеревод пока не подготовлен
Аннотация
This 14-day multicenter study assessed CT3 CGM accuracy and safety in 71 adults with diabetes. Key results: overall MARD 9.1% (median 7.1%), 82.4%-98.2% met %15/15-%30/30 criteria, 99.7% values in CEG zones A/B. CT3 showed strong performance, suitable for reliable 14-day wear.
Для цитирования: Zhou Lingli, Li Yufeng, Zhang Xingguang, Ling Hongwei, Lu Yibing, Ji Linong. A Prospective Multicenter Assessment of the Accuracy and Safety of the Yuwell CT3 Real-Time Continuous Glucose Monitoring System in Patients With Diabetes Over 14 Days.. Journal of diabetes. 2026. DOI: 10.1111/1753-0407.70246.