Clinical performance of a new point-of-care cardiac troponin I test.

Christ Michael, Geier Felicitas, Blaschke Sabine, Giannitsis Evangelos, Khellaf Mehdi, Mair Johannes, Pariente David, Scharnhorst Volkher, Semjonow Veronique, Hausfater Pierre

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

We evaluated the clinical performance of the Minicare cardiac troponin-I (cTnI), a new point-of-care (POC) cTnI test for the diagnosis of acute myocardial infarction (AMI) in a prospective, multicentre study (ISRCTN77371338).

METHODS

Of 474 patients (≥18 years) admitted to an emergency department (ED) or chest pain unit (CPU) with symptoms suggestive of acute coronary syndrome (ACS; ≤12 h from symptom onset), 465 were eligible. Minicare cTnI was tested immediately, 3 h and 6 h after presentation. AMI diagnoses were adjudicated independently based on current guidelines.

RESULTS

The diagnostic performance of the Minicare cTnI test at 3 h was similar for whole blood and in plasma: sensitivity 0.92 vs. 0.90; specificity 0.91 vs. 0.90; positive predictive value (PPV) 0.68 vs. 0.66; negative predictive value (NPV) 0.98 vs. 0.98; positive likelihood ratio (LR+) 10.18 vs. 9.41; negative likelihood ratio (LR-) 0.09 vs. 0.11. The optimal diagnostic performance was obtained at 3 h using cut-offs cTnI >43 ng/L plus cTnI change from admission ≥18.5 ng/L: sensitivity 0.90, specificity 0.96, PPV 0.81, NPV 0.98, and LR+ 21.54. The area under the receiver operating characteristics (ROC) curve for cTnI whole blood baseline value and absolute change after 3 h curve was 0.93.

CONCLUSIONS

These data support the clinical usefulness of Minicare cTnI within a 0 h/3 h-blood sampling protocol supported by current guidelines for the evaluation of suspected ACS.

Для цитирования: Christ Michael, Geier Felicitas, Blaschke Sabine, Giannitsis Evangelos, Khellaf Mehdi, Mair Johannes, Pariente David, Scharnhorst Volkher, Semjonow Veronique, Hausfater Pierre. Clinical performance of a new point-of-care cardiac troponin I test.. Clinical chemistry and laboratory medicine. 2018. DOI: 10.1515/cclm-2017-0693.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.