CD4 Lymphocyte Enumeration and Hemoglobin Assessment Aid for Priority Decisions: A Multisite Evaluation of the BD FACSPresto™ System.

Thakar Madhuri, Angira Francis, Pattanapanyasat Kovit, Wu Alan H B, O'Gorman Maurice, Zeng Hui, Qu Chenxue, Mahajan Bharati, Sukapirom Kasama, Chen Danying, Hao Yu, Gong Yan, Indig Monika De Arruda, Graminske Sharon, Orta Diana, d'Empaire Nicole, Lu Beverly, Omana-Zapata Imelda, Zeh Clement

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The BD FACSPresto™ system uses capillary and venous blood to measure CD4 absolute counts (CD4), %CD4 in lymphocytes, and hemoglobin (Hb) in approximately 25 minutes. CD4 cell count is used with portable CD4 counters in resource-limited settings to manage HIV/AIDS patients. A method comparison was performed using capillary and venous samples from seven clinical laboratories in five countries. The BD FACSPresto system was assessed for variability between laboratory, instrument/operators, cartridge lots and within-run at four sites.

METHODS

Samples were collected under approved voluntary consent. EDTA-anticoagulated venous samples were tested for CD4 and %CD4 T cells using the gold-standard BD FACSCalibur™ system, and for Hb, using the Sysmex® KX-21N™ analyzer. Venous and capillary samples were tested on the BD FACSPresto system. Matched data was analyzed for bias (Deming linear regression and Bland-Altman methods), and for concordance around the clinical decision point. The coefficient of variation was estimated per site, instrument/operator, cartridge-lot and between-runs.

RESULTS

For method comparison, 93% of the 720 samples were from HIV-positive and 7% from HIV-negative or normal subjects. CD4 and %CD4 T cells venous and capillary results gave slopes within 0.96-1.05 and R2 ≥0.96; Hb slopes were ≥1.00 and R2 ≥0.89. Variability across sites/operators gave %CV <5.8% for CD4 counts, <1.9% for %CD4 and <3.2% for Hb. The total %CV was <7.7% across instrument/cartridge lot.

CONCLUSION

The BD FACSPresto system provides accurate, reliable, precise CD4/%CD4/Hb results compared to gold-standard methods, irrespective of venous or capillary blood sampling. The data showed good agreement between the BD FACSPresto, BD FACSCalibur and Sysmex systems.

Для цитирования: Thakar Madhuri, Angira Francis, Pattanapanyasat Kovit, Wu Alan H B, O'Gorman Maurice, Zeng Hui, Qu Chenxue, Mahajan Bharati, Sukapirom Kasama, Chen Danying, Hao Yu, Gong Yan, Indig Monika De Arruda, Graminske Sharon, Orta Diana, d'Empaire Nicole, Lu Beverly, Omana-Zapata Imelda, Zeh Clement. CD4 Lymphocyte Enumeration and Hemoglobin Assessment Aid for Priority Decisions: A Multisite Evaluation of the BD FACSPresto™ System.. The open AIDS journal. 2017. DOI: 10.2174/1874613601711010076.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.