Complete Endovascular Reconstruction of the Canine Ascending Aorta, Aortic Arch, and Supra-Aortic Vessels by Implanting a New Unibody Outer Double-Branched Stent-Graft.

Li Weixiao, Li Qiang, Li Jun, Cui Mingzhe, Xu Rutao, Zhai Shuiting, Li Tianxiao, Chen Jiangbo, Zhao Wenli

СтатьяОценочное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: We evaluated the feasibility and safety of using a new unibody outer double-branched stent-graft system to reconstruct the canine ascending aorta, aortic arch, and supra-aortic vessels.

MATERIALS AND

METHODS

The outer-branched stent-graft was a unibody design. The branched stent-graft consisted of a main stent-graft and 2 branches. The introducer system included a tri-channel catheter, 2 detachable sleeves, a front fixing device, a constraining wire, and a curved outer sheath. The branched stent-graft was loaded into the introducer system. Ten adult mongrel dogs underwent general anesthesia, and the branched stent-grafts were deployed into the canine ascending aorta, aortic arch, and supra-aortic vessels by the introducer system. All animals were followed up for 3 months. At the end of the follow-up period, computed tomographic angiography (CTA) was performed to observe the patency of the branched stent-grafts.

RESULTS

The mean operation time was 142.7±13.7 minutes. The mean fluoroscopy time was 20.73±2.22 minutes. The mean dosage of contrast agent was 95.9±8.7 mL. During the operation, the tri-channel catheters successfully paralleled the wires in the aorta. All 10 branched stent-grafts were successfully implanted into the canine ascending aorta and aortic arch. There were no symptoms of cerebral embolization and no incision infection during the follow-up period. Computed tomographic angiography and specimens showed that the branched stent-grafts and native vessels were patent, the inner surfaces of the branched stent-grafts were covered by neointima, and there was no retrograde aortic dissection in the ascending aorta.

CONCLUSIONS

This animal research demonstrated that the unibody outer double-branched stent-graft system could be applied to reconstruct the canine ascending aorta, aortic arch, and supra-aortic vessels.

CLINICAL IMPACT: Thoracic endovascular aortic repair has been the main treatment method for aortic aneurysms or dissections involving the descending thoracic aorta. However, the aortic arch and ascending aorta remain the last segments of the aorta without a validated and routinely used endovascular option. In this research, we designed a new unibody outer branched stent-graft system to reconstruct the distal ascending aorta, aortic arch and supra-aortic vessels. The unibody outer branched stent-graft system could be applied to treat aortic pathologies which involve the middle and distal proximal ascending aorta, aortic arch and proximal descending aorta.

Для цитирования: Li Weixiao, Li Qiang, Li Jun, Cui Mingzhe, Xu Rutao, Zhai Shuiting, Li Tianxiao, Chen Jiangbo, Zhao Wenli. Complete Endovascular Reconstruction of the Canine Ascending Aorta, Aortic Arch, and Supra-Aortic Vessels by Implanting a New Unibody Outer Double-Branched Stent-Graft.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2022. DOI: 10.1177/15266028221139194.