MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1089/dia.2017.0163

Медицинская библиотека · Diabetes technology & therapeutics · 2017

Development of an Automatic Puncturing and Sampling System for a Self-Monitoring Blood Glucose Device.

Abe-Doi Mari, Oe Makoto, Murayama Ryoko, Takahashi Mami, Zushi Yasunobu, Tanabe Hidenori, Takamoto Iseki, Suzuki Ryo, Yamauchi Toshimasa, Kadowaki Takashi, Komiyama Chieko, Sanada Hiromi

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Performing self-monitoring of blood glucose (SMBG) is a clinical challenge for elderly people with low dexterity. An all-in-one-type SMBG device that can simply and easily puncture and monitor would be useful for them. We developed an automatic skin-puncturing and blood-sampling (APS) system for introducing of an all-in-one-type SMBG device. The aims of this study were to determine if the developed APS system, which has automatic puncturing, squeezing, and application functions, could provide sufficient blood sample volumes for SMBG and to determine the factors associated with failure in the use of the system by adult volunteers.

METHODS

We investigated the success rate of obtaining a 0.8-μL sample volume using the APS system and determined the factors associated with failure in 140 adult volunteers. The participant characteristics, induration of puncturing sites, and states of finger grip conditions were evaluated as factors of a puncturing failure. The participant characteristics, skin hydration, states of finger grip, skin elasticity of the finger pad, and blood flow were evaluated as factors of a squeezing failure.

RESULTS

The success rate was 61.9%. Puncturing failure was 21.6%, and squeezing failure was 16.5%. Automatic puncturing factors associated with failure were male sex, larger finger diameter, and thicker finger pad. The only squeezing failure factor was lower peripheral skin temperature.

CONCLUSIONS

Improvement of the finger station groove shape to prevent ischemia and the squeezing angle would be useful developments of the all-in-one-type SMBG device for elderly people with decreased dexterity.

Для цитирования: Abe-Doi Mari, Oe Makoto, Murayama Ryoko, Takahashi Mami, Zushi Yasunobu, Tanabe Hidenori, Takamoto Iseki, Suzuki Ryo, Yamauchi Toshimasa, Kadowaki Takashi, Komiyama Chieko, Sanada Hiromi. Development of an Automatic Puncturing and Sampling System for a Self-Monitoring Blood Glucose Device.. Diabetes technology & therapeutics. 2017. DOI: 10.1089/dia.2017.0163.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.