PURPOSE: To report midterm outcomes of the Nellix Endovascular Aneurysm Sealing (EVAS) System in the treatment of abdominal aortic aneurysm (AAA).

METHODS

Between September 2013 and July 2014, 64 AAA patients (mean age 76.6±6.8 years; 61 men) were treated with the EVAS system at 2 centers (only procedures performed at least 12 months prior to the analysis were included). Most patients were treated for a stable AAA, while 1 patient was treated for a ruptured aneurysm. Mean aneurysm diameter was 57.3±9.3 mm. The proximal neck measured a mean 21.5±3.3 mm in diameter and 27.0±12.1 mm long; the neck angle was 16.9°±19.3°. Eleven (17.2%) patients were treated outside the instructions for use (IFU).

RESULTS

Technical success was achieved in 63 (98.4%) of 64 patients; 1 type Ia endoleak was treated intraoperatively. One (1.6%) aneurysm-related death occurred at 4 months due to a secondary aortoenteric fistula. Overall, endoleaks occurred in 3 (4.7%) patients (2 type Ia, 1 type II). The estimated rates for 18-month overall survival, freedom from aneurysm-related death, and freedom from secondary interventions were 92.7%, 98.4%, and 95.0%, respectively. Patients treated outside the IFU had a significantly higher incidence of device-related complications (p=0.03).

CONCLUSION

The use of the Nellix device in everyday clinical practice is safe and offers promising midterm results. The risk of secondary aortoenteric fistula requires further analysis. Longer follow-up is needed to assess the actual efficacy of the device, although the risk of migration with late endoleak seems low.

Для цитирования: Silingardi Roberto, Coppi Giovanni, Ferrero Emanuele, Lauricella Antonio, Psacharopulo Daniele, Saitta Giuseppe, Viazzo Andrea, Ferri Michelangelo. Midterm Outcomes of the Nellix Endovascular Aneurysm Sealing System: A Dual-Center Experience.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2016. DOI: 10.1177/1526602816656355.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.