Computer modeling and validation testing for glenoid component rotation and optimal glenoid screw angles for reverse shoulder arthroplasty in an Asian population.

Cheng Shun Sing Martin, Yung Colin Shing-Yat, Wong Samuel De Hoi, Yip Christopher Chun Hei, Khoo Issac Jun Ren, Wong Tsoi Wan Karen, Fang Christian

СтатьяВалидационное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Good initial fixation of glenoid component for reverse total shoulder arthroplasty (RTSA) relies on component placement and screw purchase in the scapula bone. This is especially difficult in an Asian population with small glenoid geometry. Optimal glenoid component roll angle and screw angulation to achieve the longest screws for best fixation has not been defined in the current literature.

METHODS

Computer 3D modelling of 133 scapulas with RTSA performed were analyzed to determine patient specific optimal glenoid roll angle (GRA) for the longest bi-cortical screws attainable. The cranial-caudal angle (CCA), anterior-posterior angle (APA) and lengths for the superior and inferior screws were measured. Validation testing using calculated average (CA) angles and rounded average (RA) angles to the nearest 5 degree were recomputed for each case to determine the bi-cortical screw lengths achievable. The CA and RA screw lengths were compared against patient specific modelling using paired-sample t-tests.

RESULTS

Average GRA was - 1.6°, almost perpendicular to the long axis of the glenoid and achieves an average bi-cortical screw length of 51.3 mm and 45.5 mm for the superior and inferior screws respectively. The CCA and APA were 9.1° cranial and 6.5° posterior for the superior screw and screw angulation of 11.2° caudal and 0.7° anterior for the inferior screw. Validation testing shows statistically shorter screw lengths in the CA and RA models compared to patient specific modelling (p < 0.01).

CONCLUSION

Validation testing with average angles for GRA, CCA and APA demonstrates strong patient heterogeneity and anatomical variation. Despite this, screw lengths attainable in the RA group were > 38 mm with good safety profile. Surgeons may consider the additional use of navigation-assisted, or 3D printed patient specific instrumentation to optimize baseplate and screw configuration for RTSA.

Для цитирования: Cheng Shun Sing Martin, Yung Colin Shing-Yat, Wong Samuel De Hoi, Yip Christopher Chun Hei, Khoo Issac Jun Ren, Wong Tsoi Wan Karen, Fang Christian. Computer modeling and validation testing for glenoid component rotation and optimal glenoid screw angles for reverse shoulder arthroplasty in an Asian population.. International orthopaedics. 2024. DOI: 10.1007/s00264-024-06340-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.