OBJECTIVES

To evaluate the feasibility and safety of the fourth generation WATCHMAN device.

BACKGROUND

The WATCHMAN left atrial appendage (LAA) closure device has been shown to be non-inferior to conventional oral anticoagulation with warfarin for stroke prevention in patients with non-valvular atrial fibrillation. A new (fourth) generation of the WATCHMAN device was designed to facilitate easier delivery and improve safety.

METHODS

We conducted a prospective, non-randomized study of LAA closure with use of the 4(th) generation WATCHMAN device in 36 patients with non-valvular atrial fibrillation. Follow-up was performed at 45 days, 6 months and 12 months following implantation.

RESULTS

The mean age of the population was 73 ± 6 years and 67% were male. The mean CHADS2 and CHA2 DS2-VASc Score were 2.5 and 4.5, respectively. The device was implanted successfully in 34/36 (94%). No serious hospital complications related to the device or procedure occurred. During 12 month follow-up, three ischemic strokes (8%), one hemorrhagic stroke (3%), one TIA (3%) and three deaths (9%) occurred. Device-associated thrombus formation was detected in one patient (3%) during six-month follow-up and was treated successfully with low molecular weight heparin.

CONCLUSIONS

LAA closure with the 4(th) generation WATCHMAN device is feasible and safe. New technical features may contribute to a better performance of the occluder, particularly regarding the risk of pericardial effusions.

Для цитирования: Ledwoch Jakob, Franke Jennifer, Gonzaga Maik, Bertog Stefan, Braut Annkathrin, Hofmann Ilona, Vaskelyte Laura, Gafoor Sameer, Reddy Vivek, Sievert Horst. Left atrial appendage closure: First in man with the 4th generation WATCHMAN device.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2015. DOI: 10.1002/ccd.26238.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.