Lost and found: an unusual late complication of the Angio-Seal closure device.

Andreotti Felicita, Lavorgna Alberto, Coluzzi Giulio, Vincenzoni Claudio, Rebuzzi Antonio G, Porto Italo, Crea Filippo

Клинический случайПисьмо в редакциюПеревод пока не подготовлен

Аннотация

We report an unusual case of breakage and embolisation of a femoral access closure device (Angio-Seal), which caused acute leg ischemia after percutaneous coronary intervention in a diabetic patient with documented peripheral vascular disease. Embolectomy from the superficial femoral artery revealed the polymer anchor split in two parts and embedded within a large (25x5 mm) thrombus.