MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.12605

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2015

The effect of ICD programming on inappropriate and appropriate ICD Therapies in ischemic and nonischemic cardiomyopathy: the MADIT-RIT trial.

Sedláček Kamil, Ruwald Anne-Christine, Kutyifa Valentina, McNitt Scott, Thomsen Poul Erik Bloch, Klein Helmut, Stockburger Martin, Wichterle Dan, Merkely Bela, DE LA Concha Joaquin Fernandez, Swissa Moshe, Zareba Wojciech, Moss Arthur J, Kautzner Josef, Ruwald Martin H, MADIT-RIT Investigators

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

The MADIT-RIT trial demonstrated reduction of inappropriate and appropriate ICD therapies and mortality by high-rate cut-off and 60-second-delayed VT therapy ICD programming in patients with a primary prophylactic ICD indication. The aim of this analysis was to study effects of MADIT-RIT ICD programming in patients with ischemic and nonischemic cardiomyopathy.

METHODS AND

RESULTS

First and total occurrences of both inappropriate and appropriate ICD therapies were analyzed by multivariate Cox models in 791 (53%) patients with ischemic and 707 (47%) patients with nonischemic cardiomyopathy. Patients with ischemic and nonischemic cardiomyopathy had similar incidence of first inappropriate (9% and 11%, P = 0.21) and first appropriate ICD therapy (11.6% and 14.1%, P = 0.15). Patients with ischemic cardiomyopathy had higher mortality rate (6.1% vs. 3.3%, P = 0.01). MADIT-RIT high-rate cut-off (arm B) and delayed VT therapy ICD programming (arm C) compared with conventional (arm A) ICD programming were associated with a significant risk reduction of first inappropriate and appropriate ICD therapy in patients with ischemic and nonischemic cardiomyopathy (HR range 0.11-0.34, P < 0.001 for all comparisons). Occurrence of total inappropriate and appropriate ICD therapies was significantly reduced by high-rate cut-off ICD programming and delayed VT therapy ICD programming in both ischemic and nonischemic cardiomyopathy patients.

CONCLUSION

High-rate cut-off and delayed VT therapy ICD programming are associated with significant reduction in first and total inappropriate and appropriate ICD therapy in patients with ischemic and nonischemic cardiomyopathy.

Для цитирования: Sedláček Kamil, Ruwald Anne-Christine, Kutyifa Valentina, McNitt Scott, Thomsen Poul Erik Bloch, Klein Helmut, Stockburger Martin, Wichterle Dan, Merkely Bela, DE LA Concha Joaquin Fernandez, Swissa Moshe, Zareba Wojciech, Moss Arthur J, Kautzner Josef, Ruwald Martin H, MADIT-RIT Investigators. The effect of ICD programming on inappropriate and appropriate ICD Therapies in ischemic and nonischemic cardiomyopathy: the MADIT-RIT trial.. Journal of cardiovascular electrophysiology. 2015. DOI: 10.1111/jce.12605.