MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jtha.2026.02.015

Медицинская библиотека · Journal of thrombosis and haemostasis : JTH · 2026

Advancing precision diagnostics in coagulation: introduction and analytical validation of a multiplex mass spectrometry-based test for 5 coagulation factors.

Linthorst Noa A, Dirven Richard J, Romijn Fred P H T M, Mutlaq Rama, Cordfunke Robert A, Dolezal Natasja, Drijfhout Jan W, Bos Mettine H A, Eikenboom Jeroen C J, Cobbaert Christa M, van Vlijmen Bart J M, Ruhaak L Renee

СтатьяВалидационное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Circulating levels and functionality of coagulation- and fibrinolysis-related proteins are critical in bleeding and thrombosis. Traditional assays lack multiplexing and adequate standardization and require substantial plasma volumes. Quantitative protein mass spectrometry (QPMS) enables multiplexed protein quantitation, addressing clinical and research gaps.

OBJECTIVES

This study aimed to develop and validate a QPMS test for multiplexed quantitation of 5 coagulation proteins.

METHODS

A QPMS test was developed to measure plasma fibrinogen (FBG), von Willebrand factor (VWF), factor [F]VIII, FIX, and FXI. Three tryptic peptides were chosen as representatives for each protein. Stable isotope-labeled internal standards were synthesized for precise quantitation. Analytical validation assessed precision, linearity, specificity, stability, carryover, reference intervals, method comparison, and matrix effects.

RESULTS

Representative peptides for protein quantitation and identity confirmation were selected. Candidate peptides for FBG, VWF, FIX, and FXI produced sufficient signals in citrated and EDTA plasma, whereas FVIII candidate peptides did not. Validation (excluding FVIII) using plasmas from citrated healthy subjects and patients demonstrated median within-laboratory variations of 3.4% to 6.7% and a linear relationship (r = 0.990-0.998). Signals remained stable after 7 freeze/thaw cycles. A high level of agreement was shown with traditional assays (FBG: r = 0.991; VWF: r = 0.970; FIX: r = 0.926; FXI: r = 0.871). Robust relative quantitation of target proteins across different anticoagulant matrices was observed.

CONCLUSION

We developed and validated a QPMS test for multiplexed quantitation of FBG, VWF, FIX, and FXI. Enrichment is required for low-abundant FVIII. This study indicates analytical feasibility for the use of QPMS in diagnostics and justifies development of 1 QPMS test encompassing all coagulation- and fibrinolysis-related proteins.

Для цитирования: Linthorst Noa A, Dirven Richard J, Romijn Fred P H T M, Mutlaq Rama, Cordfunke Robert A, Dolezal Natasja, Drijfhout Jan W, Bos Mettine H A, Eikenboom Jeroen C J, Cobbaert Christa M, van Vlijmen Bart J M, Ruhaak L Renee. Advancing precision diagnostics in coagulation: introduction and analytical validation of a multiplex mass spectrometry-based test for 5 coagulation factors.. Journal of thrombosis and haemostasis : JTH. 2026. DOI: 10.1016/j.jtha.2026.02.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.