DOI 10.1016/j.clinbiochem.2005.11.009

Медицинская библиотека · Clinical biochemistry · 2005

Rapid micro-scale assay for homocysteine by liquid chromatography-tandem mass spectrometry.

Kuhn Joachim, Götting Christian, Kleesiek Knut

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Increased total homocysteine (tHcy) level is an independent risk factor for atherosclerosis and cardiovascular disease. Here, we describe a rapid tHcy micro-scale assay.

METHODS

We developed an easy sample preparation and rapid liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the measurement of tHcy in human serum and plasma, respectively.

RESULTS

The tHcy assay was linear up to 61.6 micromol/L, the detection limit was 1.0 micromol/L, the lower limit of quantification was 1.8 micromol/L, the imprecision in the range of 9.30-25.9 micromol/L was less than 5.9% and the analytical recovery was 94.7 +/- 6.3%.

CONCLUSIONS

Our novel tHcy LC-MS/MS assay is a quick, precise and robust method for tHcy determination in routine diagnostics.

Для цитирования: Kuhn Joachim, Götting Christian, Kleesiek Knut. Rapid micro-scale assay for homocysteine by liquid chromatography-tandem mass spectrometry.. Clinical biochemistry. 2005. DOI: 10.1016/j.clinbiochem.2005.11.009.