Yoo Albert J, Geyik Serdar, Froehler Michael T, Maurer Christoph Johannes, Kass-Hout Tareq, Zaidat Osama O, Nogueira Raul G, Hanel Ricardo A, Pierot Laurent, Spelle Laurent, Lopes Demetrius, Hassan Ameer, Širvinskas Audrius, Lin Eugene, Ribo Marc, Blasco Jordi, Taqi Muhammad Asif, Badruddin Aamir, Siddiqui Adnan H, Miller Timothy R, Hussain Shazam M, Haussen Diogo C, Woodward Keith, Groden Christoph, Consoli Arturo, Chaudry M Imran, Ramsey Christian, Maud Alberto, Bentley Joshua, Bajrami Arsida, Sahnoun Maher, Fiehler Jens, Gupta Rishi

Аннотация

BACKGROUND: Challenges to revascularization of large vessel occlusions (LVOs) persist. Current stent retrievers have limited effectiveness for removing organized thrombi. The NeVa device is a novel stent retriever designed to capture organized thrombi within the scaffold during retrieval.

OBJECTIVE: To evaluate the safety and effectiveness of revascularization of acute LVOs with the NeVa device.

METHODS: Prospective, international, multicenter, single-arm, Investigational Device Exemption study to evaluate the performance of the NeVa device in recanalizing LVOs including internal carotid artery, M1/M2 middle cerebral artery, and vertebrobasilar arteries, within 8 hours of onset. Primary endpoint was rate of expanded Treatment in Cerebral Ischemia (eTICI) score 2b-3 within 3 NeVa passes, tested for non-inferiority against a performance goal of 72% with a -10% margin. Additional endpoints included first pass success and 90-day modified Rankin Scale (mRS) score 0-2. Primary composite safety endpoint was 90-day mortality and/or 24-hour symptomatic intracranial hemorrhage (sICH).

RESULTS: From April 2021 to April 2022, 139 subjects were enrolled at 25 centers. Median National Institutes of Health Stroke Scale (NIHSS) score was 16 (IQR 12-20). In the primary analysis population (n=107), eTICI 2b-3 within 3 NeVa passes occurred in 90.7% (97/107; non-inferiority P<0.0001; post hoc superiority P<0.0001). First pass eTICI 2b-3 was observed in 73.8% (79/107), with first pass eTICI 2b67-3 in 69.2% (74/107) and eTICI 2c-3 in 48.6% (52/107). Median number of passes was 1 (IQR 1-2). Final eTICI 2b-3 rate was 99.1% (106/107); final eTICI 2b67-3 rate was 91.6% (98/107); final eTICI 2c-3 rate was 72.9% (78/107). Good outcome (90-day mRS score 0-2) was seen in 65.1% (69/106). Mortality was 9.4% (13/138) with sICH in 5.0% (7/139).

CONCLUSIONS: The NeVa device is highly effective and safe for revascularization of LVO strokes and demonstrates superior first pass success compared with a predicate performance goal.

TRIAL REGISTRATION NUMBER: NCT04514562.