DOI 10.3760/cma.j.issn.1001-9391.2019.09.014
[Evaluation of GeneXpert MTB/RIF and BACTEC-MGIT 960 for the detection of tuberculosis among pneumoconiosis-associated tuberculosis patients].
Аннотация
Objective
To evaluate the performence of GeneXpert MTB/RIF and BACTEC-MGIT 960 on detecting Mycobacterium tuberculosis and rifampicin resistance for pneumoconiosis-associated tuberculosis patients.
Methods
The recruited 133 suspected active pneumoconiosis-associated tuberculosis hospitalized cases, morning sputum samples were collected to do modified L-J culture, conventional proportion method drug susceptibility test, GeneXpert MTB/RIF and BACTEC-MGIT 960. Analyze the sensitivity and specificity of the 133 sputum from patients, the positive rates of patients with tuberculosis in GeneXpert MTB/RIF test, BACTEC-MGIT 960 and modified L-J culture were 37.59%, 34.59% and 30.08% respectively. There was no significant difference among the three tests respectively (P>0.05) . According to the modified L-J culture, the sensitivity of GeneXpert MTB/RIF and BACTEC-MGIT 960 in detecting tuberculosis were 92.5% and 95.0% respectively, and specificity in rifampicin resistance were 86.0% and 91.4% respectively. There was no significant difference between GeneXpert MTB/RIF and BACTEC-MGIT 960 (P>0.05) . According to conventional proportion method drug susceptibility test, the sensitivity of GeneXpert MTB/RIF and BACTEC-MGIT 960 in detecting rifampicin resistance were 90.0% and 100%, and specificity were 92.6% and 96.4%. There was no significant difference between GeneXpert MTB/RIF and BACTEC-MGIT 960 (P>0.05) .
Conclusion
The GeneXpert MTB/RIF has good performence of detecting tuberculosis and rifampicin resistance. It has good application value among pneumoconiosis-associated tuberculosis patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.