DOI 10.1093/europace/euaf170

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2025

Preventing smartphone induction of magnet mode in cardiac implantable electronic devices.

Wegner Felix K, Korthals Dennis, Kreimer Fabienne, Wolfes Julian, Ellermann Christian, Martinovic Moritz, Kittl Christoph, Frommeyer Gerrit, Eckardt Lars

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Wireless charging capable smartphones may interact with cardiac implantable electronic devices (CIED). We hypothesized that magnetic shielding with a steel plate placed at the back of a smartphone may prevent interaction.

METHODS AND

RESULTS

Sixteen CIED (6 pacemakers and 10 implantable cardioverters/defibrillators) from all manufacturers were consecutively implanted in a subcutaneous and submuscular location in an isolated porcine thorax and connected to an interactive heart simulator. Two smartphones (Apple iPhone 14 and Google Pixel 8 Pro) were placed on top of the implantation site, and signs of magnet mode induction were recorded. The Apple iPhone induced magnet mode in seven of 16 (44%) subcutaneously placed CIED. Placed in a magnetic phone case, the Apple iPhone induced magnet mode in six of 16 (38%) devices. Magnet mode induction was successfully prevented in all cases by placing a 1 mm thin steel plate at the back of the smartphone. The Google Pixel did not induce magnet mode in any of the 16 devices, even with a magnetic case. Submuscular CIED placement resulted in no occurrence of magnet mode induction by any smartphone. All devices were still able to be interrogated and responded to a CIED magnet.

CONCLUSION

The Apple iPhone induces magnet mode in close to half of subcutaneously placed CIED due to its MagSafe magnet. Neither a smartphone utilizing the Qi wireless charging standard nor magnetic smartphone cases confer the same risk. Submuscular CIED placement or magnetic shielding with a steel plate prevents magnet mode induction. The future Qi2.0 standard may exacerbate interaction risks.

Для цитирования: Wegner Felix K, Korthals Dennis, Kreimer Fabienne, Wolfes Julian, Ellermann Christian, Martinovic Moritz, Kittl Christoph, Frommeyer Gerrit, Eckardt Lars. Preventing smartphone induction of magnet mode in cardiac implantable electronic devices.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2025. DOI: 10.1093/europace/euaf170.