MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.echo.2011.03.009

Медицинская библиотека · Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography · 2011

Left atrial volume by real-time three-dimensional echocardiography: validation by 64-slice multidetector computed tomography.

Miyasaka Yoko, Tsujimoto Satoshi, Maeba Hirofumi, Yuasa Fumio, Takehana Kazuya, Dote Kinuko, Iwasaka Toshiji

СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Left atrial (LA) enlargement has been acknowledged as a significant predictor of cardiovascular morbidity and mortality.

METHODS

To evaluate the accuracy of two-dimensional and three-dimensional echocardiography for determining LA volume, LA volume measurements by echocardiography were compared with those measured by 64-slice multidetector computed tomography (MDCT) as a reference standard.

RESULTS

Fifty-seven consecutive patients (mean age, 66 ± 11 years; 59% men) referred to echocardiography and MDCT on the same day were prospectively evaluated. LA volume by three-dimensional echocardiography was correlated closely with that by MDCT (r = 0.95, P < .0001), with 8% underestimation. LA volume by two-dimensional echocardiography was correlated less well with that measured by MDCT (r = 0.86, P < .0001) and consistently underestimated LA volume by 19%, particularly as the left atrium enlarged.

CONCLUSIONS

LA volume assessment by three-dimensional echocardiography was correlated closely with that measured by MDCT, albeit with an 8% underestimation. Three-dimensional echocardiography is a feasible noninvasive method to evaluate LA volume.

Для цитирования: Miyasaka Yoko, Tsujimoto Satoshi, Maeba Hirofumi, Yuasa Fumio, Takehana Kazuya, Dote Kinuko, Iwasaka Toshiji. Left atrial volume by real-time three-dimensional echocardiography: validation by 64-slice multidetector computed tomography.. Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography. 2011. DOI: 10.1016/j.echo.2011.03.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.