m+Medelevator.RU
← Поиск публикаций
Thorax · 2000

Comparison of two in-laboratory titration methods to determine effective pressure levels in patients with obstructive sleep apnoea.

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Bureau M P, Sériès F

Аннотация

BACKGROUND: Treatment of the sleep apnoea/hypopnoea syndrome with nasal continuous positive airway pressure (CPAP) conventionally requires a titration procedure in a sleep laboratory. The upper airway has a hysteresis phenomenon which accounts for a decrease in the effective positive pressure level once an initial effective pressure setting has been reached. The aim of this study was to quantify the difference in the initial and final effective pressure settings when the titration sleep study takes into account these possible changes in the effective pressure level.

METHODS: Eighty five patients completed the study. During a titration sleep study the pressure was increased by 1 cm H2O in a stepwise fashion until obstructive respiratory events disappeared (effective pressure 1, Peff(1)). The pressure level was then decreased by increments of 1 cm H2O until breathing abnormalities reappeared. At this time pressure was re-increased by increments of 1 cm H(2)O to normalise breathing (Peff(2)).

RESULTS: The mean (SD) value of Peff(1) was 9.5 (2.6) cm H2O. The pressure was then reduced during 0.5 (0.6) hours to reach the minimal pressure (7.0 (2.0) cm H2O). The pressure obtained after a downward titration had to be re-increased in 79 patients, the Peff(2) level being significantly lower than Peff(1) (8.9 (2.8) cm H2O, p = 0.0002), mean difference 0.6 (1.5) cm H2O (95% confidence interval 0.29 to 0.93).

CONCLUSION: Attempts to decrease the positive pressure level during conventional determination of the effective pressure level allow a significant decrease in the pressure setting. This should be taken into account in each patient who requires an in-laboratory manual CPAP titration procedure.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты