MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1089/dia.2020.0224

Медицинская библиотека · Diabetes technology & therapeutics · 2020

Compatibility and Safety of Ultra Rapid Lispro with Continuous Subcutaneous Insulin Infusion in Patients with Type 1 Diabetes: PRONTO-Pump Study.

Bode Bruce W, Garg Satish K, Norwood Paul, Morales Cristobal, Hardy Thomas, Liu Rong, Ignaut Debra

Клиническое исследование III фазыСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Background

Ultra rapid lispro (URLi) is a new insulin lispro formulation that has accelerated absorption and improved postprandial glucose control compared with insulin lispro (Humalog®). The compatibility and safety of URLi versus lispro were evaluated in patients with type 1 diabetes using continuous subcutaneous insulin infusion (insulin pump).

Methods

In this phase 3, double-blind, crossover study, 49 patients were randomized to two 6-week treatment periods, after a 2-week lead-in period on lispro. The primary endpoint was the rate of infusion set failures due to a pump occlusion alarm, or unexplained hyperglycemia with blood glucose >13.9 mmol/L (250 mg/dL) that did not decrease within 1 h after a correction bolus.

Results

There was no significant difference in the rate of infusion set failures between URLi and lispro (0.03 vs. 0.05 events/30 days, P = 0.375). A higher rate of premature infusion set changes was observed with URLi (1.13 vs. 0.78 events/30 days; P = 0.028), translating to one additional infusion set change approximately every 3 months. A trend toward improved glycemic control was observed with URLi treatment: Time in range 3.9-10.0 mmol/L (71-180 mg/dL) was 65.7% ± 1.3% versus 63.0% ± 1.3%. Treatment-emergent adverse events (TEAEs) were reported by 46.9% of patients on URLi treatment and 18.8% on lispro. This difference was driven by an increase in infusion site reactions-more than 90% were mild. Incidence of all other TEAEs and severe hypoglycemia was similar between treatments.

Conclusions

URLi was compatible with insulin pump use with a safety profile similar to lispro.

Для цитирования: Bode Bruce W, Garg Satish K, Norwood Paul, Morales Cristobal, Hardy Thomas, Liu Rong, Ignaut Debra. Compatibility and Safety of Ultra Rapid Lispro with Continuous Subcutaneous Insulin Infusion in Patients with Type 1 Diabetes: PRONTO-Pump Study.. Diabetes technology & therapeutics. 2020. DOI: 10.1089/dia.2020.0224.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.