JACC. Cardiovascular interventions · 2015
Treatment of Heart Failure With Associated Functional Mitral Regurgitation Using the ARTO System: Initial Results of the First-in-Human MAVERIC Trial (Mitral Valve Repair Clinical Trial).
СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rogers Jason H, Thomas Martyn, Morice Marie-Claude, Narbute Inga, Zabunova Milana, Hovasse Thomas, Poupineau Mathieu, Rudzitis Ainars, Kamzola Ginta, Zvaigzne Ligita, Greene Samantha, Erglis Andrejs
Аннотация
OBJECTIVES: MAVERIC (Mitral Valve Repair Clinical Trial) reports the safety and efficacy of the ARTO system in patients with symptomatic heart failure and functional mitral regurgitation (FMR).
BACKGROUND: The ARTO system percutaneously modifies the mitral annulus to improve leaflet coaptation in FMR.
METHODS: The MAVERIC trial is a prospective, nonrandomized first-in-human study. Key inclusion criteria were systolic heart failure New York Heart Association functional classes II to IV, FMR grade ≥2+, left ventricular (LV) ejection fraction ≤40%, LV end-diastolic diameter >50 mm and ≤75 mm. Exclusion criteria were clinical variables that precluded feasibility of the ARTO procedure. Primary outcomes were safety (30-day major adverse events) and efficacy (MR reduction, LV volumes, and functional status).
RESULTS: Eleven patients received the ARTO system, and there were no procedural adverse events. From baseline to 30 days, there were meaningful improvements. Effective regurgitant orifice area decreased from 30.3 ± 11.1 mm(2) to 13.5 ± 7.1 mm(2) and regurgitant volumes from 45.4 ± 15.0 ml to 19.5 ± 10.2 ml. LV end-systolic volume index improved from 77.5 ± 24.3 ml/m(2) to 68.5 ± 21.4 ml/m(2), and LV end-diastolic volume index 118.7 ± 28.6 ml/m(2) to 103.9 ± 21.2 ml/m(2). Mitral annular anteroposterior diameter decreased from 45.0 ± 3.3 mm to 38.7 ± 3.0 mm. Functional status was 81.8% New York Heart Association functional class III/IV improving to 54.6% functional class I/II. At 30 days, there were 2 adverse events: 1 pericardial effusion requiring surgical drainage; and 1 asymptomatic device dislodgement.
CONCLUSIONS: The ARTO system is a novel transcatheter device that can be used safely with meaningful efficacy in the treatment of FMR. (Mitral Valve Repair Clinical Trial [MAVERIC]; NCT02302872).
BACKGROUND: The ARTO system percutaneously modifies the mitral annulus to improve leaflet coaptation in FMR.
METHODS: The MAVERIC trial is a prospective, nonrandomized first-in-human study. Key inclusion criteria were systolic heart failure New York Heart Association functional classes II to IV, FMR grade ≥2+, left ventricular (LV) ejection fraction ≤40%, LV end-diastolic diameter >50 mm and ≤75 mm. Exclusion criteria were clinical variables that precluded feasibility of the ARTO procedure. Primary outcomes were safety (30-day major adverse events) and efficacy (MR reduction, LV volumes, and functional status).
RESULTS: Eleven patients received the ARTO system, and there were no procedural adverse events. From baseline to 30 days, there were meaningful improvements. Effective regurgitant orifice area decreased from 30.3 ± 11.1 mm(2) to 13.5 ± 7.1 mm(2) and regurgitant volumes from 45.4 ± 15.0 ml to 19.5 ± 10.2 ml. LV end-systolic volume index improved from 77.5 ± 24.3 ml/m(2) to 68.5 ± 21.4 ml/m(2), and LV end-diastolic volume index 118.7 ± 28.6 ml/m(2) to 103.9 ± 21.2 ml/m(2). Mitral annular anteroposterior diameter decreased from 45.0 ± 3.3 mm to 38.7 ± 3.0 mm. Functional status was 81.8% New York Heart Association functional class III/IV improving to 54.6% functional class I/II. At 30 days, there were 2 adverse events: 1 pericardial effusion requiring surgical drainage; and 1 asymptomatic device dislodgement.
CONCLUSIONS: The ARTO system is a novel transcatheter device that can be used safely with meaningful efficacy in the treatment of FMR. (Mitral Valve Repair Clinical Trial [MAVERIC]; NCT02302872).