Pressure monitoring inside Meckel's cave during percutaneous microcompression of gasserian ganglion.
Аннотация
During percutaneous microcompression of the gasserian ganglion for the relief of trigeminal neuralgia, a computerized technique for monitoring the pressure inside Meckel's cave was employed in 22 patients. A dedicated transducer connected to a computer records the balloon inflation pressure. Its variations are discernible within tenths of a bar and are plotted in relation to time. The intraoperative pressure inside Meckel's cave is from 0.9 to 2.4 bars. When pressure was low, there was recurrence of pain. The highest values of pressure (1.9-2.4 bars) were observed in most of the patients suffering from untoward side effects. The clinical results seem to be influenced by the level of the intraoperative intracavitary pressure.