Evaluation of reticulocyte hemoglobin content (RETIC-HGB) for the diagnosis of iron-limited erythropoiesis in cats.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Keiner Miriam, Fuchs Jannika, Bauer Natali, Moritz Andreas
Аннотация
BACKGROUND: Decreased reticulocyte hemoglobin content (CHr) (Siemens ADVIA 2120) reflects iron-limited erythropoiesis (ILE). RETIC-HGB (IDEXX ProCyte Dx) is a novel marker of ILE for veterinary use.
OBJECTIVES: We aimed to evaluate reference intervals (RIs) and the utility of RETIC-HGB and CHr in the diagnosis of feline ILE.
MATERIALS AND METHODS: RIs were established in 59 healthy cats. Intra-assay coefficients of variation (CVs) and correlations between RETIC-HGB and CHr were assessed. Two hundred and seventy-five cats were classified as having ILE or not based on low plasma iron or low transferrin saturation along with anemia and/or altered RBC indices. CHr, RETIC-HGB, and serum amyloid A (SAA) were compared between the groups. The sensitivity and specificity of RETIC-HGB and CHr to diagnose ILE were analyzed to determine the RI lower limits.
RESULTS: RIs for RETIC-HGB and CHr were 12.5-18.0 and 14.0-19.9 pg, respectively. The CV was 3% for both variables. RETIC-HGB and CHr were moderately correlated (rs = 0.59) with a bias of -1.2 picograms (pgs). Twenty of the 275 cats were classified as having ILE. Compared with non-ILE cats, ILE cats had significantly lower median RETIC-HGB (14.3 vs 15.2 pg, P = .0046) and mean CHr (14.7 vs 16.5 pg, P < .0001) values and significantly increased median SAA (44.6 vs 2.3 µg/dl, P < .0001) values. Using the lower RI limits resulted in a low sensitivity and relatively high specificity to diagnose ILE in cats.
CONCLUSIONS: ILE was characterized by decreased CHr and RETIC-HGB; however, sensitivity was low. The moderate correlation between RETIC-HGB and CHr is likely due to species differences and different methodology.
OBJECTIVES: We aimed to evaluate reference intervals (RIs) and the utility of RETIC-HGB and CHr in the diagnosis of feline ILE.
MATERIALS AND METHODS: RIs were established in 59 healthy cats. Intra-assay coefficients of variation (CVs) and correlations between RETIC-HGB and CHr were assessed. Two hundred and seventy-five cats were classified as having ILE or not based on low plasma iron or low transferrin saturation along with anemia and/or altered RBC indices. CHr, RETIC-HGB, and serum amyloid A (SAA) were compared between the groups. The sensitivity and specificity of RETIC-HGB and CHr to diagnose ILE were analyzed to determine the RI lower limits.
RESULTS: RIs for RETIC-HGB and CHr were 12.5-18.0 and 14.0-19.9 pg, respectively. The CV was 3% for both variables. RETIC-HGB and CHr were moderately correlated (rs = 0.59) with a bias of -1.2 picograms (pgs). Twenty of the 275 cats were classified as having ILE. Compared with non-ILE cats, ILE cats had significantly lower median RETIC-HGB (14.3 vs 15.2 pg, P = .0046) and mean CHr (14.7 vs 16.5 pg, P < .0001) values and significantly increased median SAA (44.6 vs 2.3 µg/dl, P < .0001) values. Using the lower RI limits resulted in a low sensitivity and relatively high specificity to diagnose ILE in cats.
CONCLUSIONS: ILE was characterized by decreased CHr and RETIC-HGB; however, sensitivity was low. The moderate correlation between RETIC-HGB and CHr is likely due to species differences and different methodology.