Analytical Validation of an Automated Point-of-Care Immunoassay for the Measurement of Feline Pancreatic Lipase Immunoreactivity Concentration (Vcheck fPL 2.0).
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wetzel Lisa, Steiner Joerg M, Cridge Harry
Аннотация
BACKGROUND: Pancreatic lipase assays are commonly utilized in the diagnostic approach to suspected pancreatitis in cats. The Vcheck fPL 2.0 assay is commercially available; however, analytical validation has yet to be reported in the peer-reviewed literature.
OBJECTIVE: To evaluate the analytic validity of a fluorescent immunoassay for the point-of-care quantification of feline pancreatic lipase immunoreactivity (Vcheck fPL 2.0 assay).
METHODS: Linearity was assessed via dilutional parallelism. Intra- and interassay variability was assessed by calculating coefficients of variation (%CV) across replicates on the same day and across multiple days, respectively. Bilirubin, hemoglobin, and intralipid (intralipid 30% IV emulsion, VWR, Radnor, PA, USA) were utilized to evaluate the potential effects of interfering substances on the assay.
RESULTS: Linearity was sub-optimal, with 12/15 diluted runs having an O/E ratio outside of the 80-120% target range. The mean %CV was 10.2% for intra-assay variability and 16.3% for interassay variability. Interfering substances had no significant effect on the results of the Vcheck fPL 2.0 assay (P = 0.888).
CONCLUSIONS: The Vcheck fPL 2.0 appears to have poor linearity and suboptimal precision, limiting the accurate interpretation of results. Additional assay optimization should be considered prior to routine clinical use.
OBJECTIVE: To evaluate the analytic validity of a fluorescent immunoassay for the point-of-care quantification of feline pancreatic lipase immunoreactivity (Vcheck fPL 2.0 assay).
METHODS: Linearity was assessed via dilutional parallelism. Intra- and interassay variability was assessed by calculating coefficients of variation (%CV) across replicates on the same day and across multiple days, respectively. Bilirubin, hemoglobin, and intralipid (intralipid 30% IV emulsion, VWR, Radnor, PA, USA) were utilized to evaluate the potential effects of interfering substances on the assay.
RESULTS: Linearity was sub-optimal, with 12/15 diluted runs having an O/E ratio outside of the 80-120% target range. The mean %CV was 10.2% for intra-assay variability and 16.3% for interassay variability. Interfering substances had no significant effect on the results of the Vcheck fPL 2.0 assay (P = 0.888).
CONCLUSIONS: The Vcheck fPL 2.0 appears to have poor linearity and suboptimal precision, limiting the accurate interpretation of results. Additional assay optimization should be considered prior to routine clinical use.