DOI 10.1055/s-2005-865689

Медицинская библиотека · The Thoracic and cardiovascular surgeon · 2005

Different tilting disc valves show similar rotation-dependent impairment in hemodynamic performance under a tilted implantation position.

Hartrumpf M, Kuehnel R U, Puchner R, Wendt M O, Pohl M, Albes J M

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Aortic annulus calcification can promote tilted implantation of mechanical valves. This study evaluates the hemodynamics of tilting disc valves under this condition.

METHODS

23 mm and 25 mm Ultracor (UC) and Medtronic-Hall-Easy-Fit (MH) valves were investigated in a pulse-duplicator under physiological conditions. Mean pressure gradient (dP(mean)), systolic energy loss (dW(sys)), effective orifice area (EOA), closure (V(Cl)), leakage (V(L)), and total regurgitation volume (V(R)) were assessed. Valves were independently positioned at five axial rotations (0 - 180 degrees , zero defined as major orifice facing the top of the "tilt-ramp") and three tilt angles (0 degrees, 10 degrees, 20 degrees) by lifting the prosthesis in the noncoronary sinus.

RESULTS

Diameter-enhanced MH valves exhibited a better systolic performance but a higher regurgitation than corresponding UC valves. Moderate tilting showed a rotation-independent increase in dP(mean) and dW(sys) and a decrease in V (R) and EOA with no fundamental differences between valve types. Further tilting caused small additional changes at 90 - 180 degrees rotation. At 0 degrees rotation, however, dramatic regurgitation occurred throughout.

CONCLUSION

Tilting worsened systolic performance regardless of valve type. It should therefore be avoided. Due to extensive regurgitation at 0 degrees rotation, this position should be corrected whenever tilting is inevitable.

Для цитирования: Hartrumpf M, Kuehnel R U, Puchner R, Wendt M O, Pohl M, Albes J M. Different tilting disc valves show similar rotation-dependent impairment in hemodynamic performance under a tilted implantation position.. The Thoracic and cardiovascular surgeon. 2005. DOI: 10.1055/s-2005-865689.