Hemodynamic performance and incidence of patient-prosthesis mismatch of the complete supraannular perimount magna bioprosthesis in the aortic position.

Botzenhardt F, Eichinger W B, Guenzinger R, Bleiziffer S, Wagner I, Bauernschmitt R, Lange R

Клиническое исследованиеСравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Complete supraannular placement and smaller stent design allow the implantation of a Perimount Magna bioprosthesis with a larger inner diameter than that of a standard Perimount. This study compares the hemodynamic performance and the incidence of patient-prosthesis mismatch (PPM) of both prostheses.

METHODS

128 patients underwent aortic valve replacement, receiving either a Magna (n = 57) or a standard (n = 71) prosthesis. Inner aortic annulus diameter was measured intraoperatively by a hegar dilator to match echocardiographically obtained results to the annulus diameter instead of matching them to labelled valve size.

RESULTS

The Magna was significantly superior with respect to mean pressure gradient and effective orifice area in patients with an annulus diameter of 22-23 mm. In patients with an annulus diameter < 22 mm or > 23 mm, there was a non-significant trend towards superior hemodynamics in the Magna group. Severe PPM (effective orifice area index < or = 0.65 cm (2)/m (2)) was present in 11.1% (Magna) vs. 42.1% (Standard) of patients with an annulus diameter < 22 mm; in 0% (Magna) vs. 13.8% (Standard) with an annulus diameter of 22-23 mm; no PPM was seen in patients with annulus diameter > 23 mm in both groups.

CONCLUSIONS

The Perimount Magna had a significantly reduced incidence of patient-prosthesis mismatch and superior hemodynamics compared to the standard Perimount.

Для цитирования: Botzenhardt F, Eichinger W B, Guenzinger R, Bleiziffer S, Wagner I, Bauernschmitt R, Lange R. Hemodynamic performance and incidence of patient-prosthesis mismatch of the complete supraannular perimount magna bioprosthesis in the aortic position.. The Thoracic and cardiovascular surgeon. 2005. DOI: 10.1055/s-2005-837678.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.