Indication and electrical performance of conventional, resynchronization, and conduction system pacing in transthyretin amyloid cardiomyopathy.

Dobner Stephan, Tawo Serlha, Noti Fabian, Wieser Fabian, Gräni Christoph, Nitsche Christian, Reichlin Tobias, Hunziker Lukas, Haeberlin Andreas

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Conduction system disorders and arrhythmias may require pacemaker (PM) or implantable cardioverter-defibrillator implantation in transthyretin amyloid cardiomyopathy (ATTR-CM). The optimal mode of pacing in ATTR-CM remains unknown.

OBJECTIVE

This study aimed to compare the electrical performance and clinical outcomes of conventional, resynchronization, and conduction system pacing (CSP) in ATTR-CM.

METHODS

In this observational study, the effects of cardiac pacing were investigated in 250 consecutive patients with ATTR-CM presenting at Inselspital Bern between June 2019 and February 2023.

RESULTS

During follow-up, 67 of 250 patients (26.8%) received a PM. Implantation of conventional single- (VVI) or dual-chamber (DDD) PMs was more common before ATTR-CM diagnosis (n = 17 of 25; 68%), and physiological pacing (cardiac resynchronization therapy [CRT]/CSP) was increasingly used thereafter (n = 24 of 42 [57.1%]). Sick sinus syndrome (n = 11 of 35 [31.4%] vs 6 of 32 [18.8%]) and higher-degree atrioventricular block (n = 20 of 35 [57.1%] vs 9 of 32 [28.1%]) were more common indications for VVI/DDD, with pursuit of a pace/ablate strategy (n = 12 of 32 [37.5%]) and heart failure (HF) (n = 5 of 32 [15.6%]) contributing significantly to implantation of CRT/CSP. QRS duration width was significantly lower with CSP (122 ms [interquartile range {IQR} 120-139]) than CRT (155 ms [IQR 141-160]) or VVI/DDD (160 ms [IQR 144-180]; P < .001). Although left ventricular/left bundle branch area pacing capture thresholds were significantly lower after CSP (0.75 V [IQR 0.5-1.2]) than CRT (1.3 V [IQR 1.0-1.6]; P = .049) at implantation, no significant differences for atrial or ventricular lead performance were observed during follow-up (P > .05). Patients referred for CRT/CSP had a higher incidence of HF (P = .014) and more common HF hospitalizations (P = .03); however, mortality was not significantly different (P = .28).

CONCLUSION

Offering stable electrical performance and improved resynchronization, the effect of CSP on clinical outcomes warrants further exploration in ATTR-CM.ClinicalTrials.gov identifier: NCT04776824.

Для цитирования: Dobner Stephan, Tawo Serlha, Noti Fabian, Wieser Fabian, Gräni Christoph, Nitsche Christian, Reichlin Tobias, Hunziker Lukas, Haeberlin Andreas. Indication and electrical performance of conventional, resynchronization, and conduction system pacing in transthyretin amyloid cardiomyopathy.. Heart rhythm O2. 2026. DOI: 10.1016/j.hroo.2026.03.026.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.