MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1089/lrb.2018.0002

Медицинская библиотека · Lymphatic research and biology · 2018

Electrode Equivalence for Use in Bioimpedance Spectroscopy Assessment of Lymphedema.

Svensson Brenda J, Dylke Elizabeth S, Ward Leigh C, Kilbreath Sharon L

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Bioimpedance spectroscopy (BIS) is commonly used in the assessment and monitoring of lymphedema. This study investigated electrodes as a source of variability that could impact the accuracy of BIS in the clinic and determined if Ag/AgCl electrocardiograph (ECG) electrodes could be used as an alternative to instrument-specific electrodes.

METHODS AND

RESULTS

Two types of Ag/AgCl electrodes were studied: instrument-specific bioimpedance electrodes (bioimpedance) and single tab ECG electrodes (cardiac). Six areas of investigation were addressed: intrinsic electrode resistance; electrode age; drive electrode position; electrode width/surface area; concordance between cardiac and bioimpedance electrodes; and mixing electrode types and batches. Participants included women (n = 26) and men (n = 8), both with (n = 4) and without lymphedema (n = 30). Resistance (R0) of the limbs was measured and used to calculate interlimb BIS ratios. Intrinsic electrode resistance varied between batches (p ≤ 0.001), with cardiac electrodes recording higher resistance. Electrode age had no impact on limb resistance (p = 0.85). Drive electrode position biased limb resistance (0.1%-2.3%) and electrode size/surface area had a small (≤1%), but significant effect on limb resistance (p ≤ 0.001). However, calculation of interlimb BIS ratios negated the impact of these as well as any effect of mixing electrode batches and types (p = 0.15-0.96). Electrode type had no impact on arm and leg resistance, or interlimb BIS ratios (p = 0.173-0.289).

CONCLUSION

Calculation of interlimb BIS ratios improves accuracy of clinical BIS. Ag/AgCl cardiac electrodes can be used as an alternative to device-specific electrodes to measure limb resistance.

Для цитирования: Svensson Brenda J, Dylke Elizabeth S, Ward Leigh C, Kilbreath Sharon L. Electrode Equivalence for Use in Bioimpedance Spectroscopy Assessment of Lymphedema.. Lymphatic research and biology. 2018. DOI: 10.1089/lrb.2018.0002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.