Robotic Transabdominal Control of the Suprahepatic, Infradiaphragmatic Vena Cava to Enable Level 3 Caval Tumor Thrombectomy: Pilot Study in a Perfused-Cadaver Model.

de Castro Abreu Andre Luis, Chopra Sameer, Azhar Raed A, Berger Andre K, Metcalfe Charles, Minetti Michael, Carey Joseph N, Ukimura Osamu, Desai Mihir, Gill Inderbir S

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To develop a robotic technique for exclusively transabdominal control of the suprahepatic, infradiaphragmatic inferior vena cava (IVC) to enable level 3 IVC tumor thrombectomy.

MATERIALS AND

METHODS

Robotic technique was developed in three fresh, perfused-model cadavers. Preoperatively, inflow (right jugular vein) and outflow (left femoral vein) cannulae were inserted and connected to a centrifugal pump to establish a 10 mmHg pressure in the IVC for the water-perfused cadaver model. Using a five-port transperitoneal robotic approach, the falciform ligament was detached from the anterior abdominal wall toward its junction with the diaphragm and tautly retracted caudally; this adequately retracted the liver caudally as well. Triangular and coronary ligaments were incised, allowing ready visualization of suprahepatic/infradiaphragmatic IVC and right/left main hepatic veins. Under direct robotic visualization, IVC was circumferentially mobilized, vessel-looped, and controlled.

RESULTS

All three robotic procedures were successfully completed transabdominally. Average robotic time to control the suprahepatic IVC was 37 minutes; in each case, the suprahepatic IVC was circumferentially controlled with a vessel-loop. There were no intraoperative complications. Length of the mobilized suprahepatic IVC measured between 2 and 3 cm. Right and left suprahepatic veins were clearly visualized in each case. Necropsy revealed no intra-abdominal/intrathoracic visceral or vascular injuries to the suprahepatic IVC, bilateral hepatic veins, or tributaries.

CONCLUSIONS

We developed a novel robotic technique for transabdominal control of the suprahepatic infradiaphragmatic IVC in a perfused human cadaver model. This approach may extend the application of advanced robotic techniques for the performance of major vena caval, hepatic, and level 3 IVC renal tumor thrombus surgery.

Для цитирования: de Castro Abreu Andre Luis, Chopra Sameer, Azhar Raed A, Berger Andre K, Metcalfe Charles, Minetti Michael, Carey Joseph N, Ukimura Osamu, Desai Mihir, Gill Inderbir S. Robotic Transabdominal Control of the Suprahepatic, Infradiaphragmatic Vena Cava to Enable Level 3 Caval Tumor Thrombectomy: Pilot Study in a Perfused-Cadaver Model.. Journal of endourology. 2015. DOI: 10.1089/end.2015.0081.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.