MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2016.01.031

Медицинская библиотека · Heart rhythm · 2016

Extraction of chronically implanted coronary sinus leads active fixation vs passive fixation leads.

Crossley George H, Sorrentino Robert A, Exner Derek V, Merliss Andrew D, Tobias Serge M, Martin David O, Augostini Ralph, Piccini Jonathan P, Schaerf Raymond, Li Shelby, Miller Clayton T, Adler Stuart W

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The Medtronic model 4195 (StarFix) left ventricular lead is an active fixation lead that provides additional support within the coronary sinus (CS) via deployable lobes. While this lead has been shown to have excellent stability within the CS, concerns about its extractability have been raised.

OBJECTIVE

The aim of this study was to compare the safety and efficacy of the extraction of the model 4195 lead vs other Medtronic CS leads in a prospective cohort study.

METHODS

Patients undergoing extraction of this and other CS leads for standard indications were prospectively enrolled and studied. The primary outcomes of interest were the removal success rates and associated complication rates. Patients were followed for a month postprocedure.

RESULTS

The overall left ventricular lead extraction success rate was 97.6% (n = 205). Among 40 patients with chronic model 4195 leads, there were 37 successful extractions (92.5%) as compared to 98.8% for the 165 non-4195 leads. However, in 2 of the 3 StarFix lead extraction failures, standard extraction techniques were not used. All 10 of the model 4195 leads that had been implanted for less than 6 months were extracted without incident.

CONCLUSION

In this largest study of CS lead extractions published to date, the overall success rate of the extraction of chronically implanted CS leads is high and the complication rate is similar in these lead models. The extraction of the model 4195 lead is clearly more challenging, but it can be accomplished in high-volume extraction centers with experienced operators. It is recommended that the model 4195 lead be extracted by experienced operators.

Для цитирования: Crossley George H, Sorrentino Robert A, Exner Derek V, Merliss Andrew D, Tobias Serge M, Martin David O, Augostini Ralph, Piccini Jonathan P, Schaerf Raymond, Li Shelby, Miller Clayton T, Adler Stuart W. Extraction of chronically implanted coronary sinus leads active fixation vs passive fixation leads.. Heart rhythm. 2016. DOI: 10.1016/j.hrthm.2016.01.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.