Misleading FT4 and FT3 Due to Immunoassay Interference From Autoantibodies.

Goettemoeller Travis, McShane Adam J, Rao Pratibha

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Immunoassays used to measure total and free thyroxine (T4 and FT4) and total and free triiodothyronine (T3 and FT3) can provide inaccurate results due to interference from endogenous autoantibodies.

CASE REPORT: A 74-year-old female treated for hypothyroidism with levothyroxine replacement had elevated thyroid stimulating hormone (TSH), FT4, and FT3. Due to concern for hyperthyroidism, levothyroxine was discontinued and further workup was initiated. A pituitary MRI revealed a microadenoma but the alpha-subunit was normal. She was given octreotide for suspected TSH secreting pituitary adenoma without improvement in her TSH, FT4, or FT3 levels. She was referred to our clinic, where inaccurate lab values for FT4 and FT3 were suspected.

RESULTS

Testing via equilibrium dialysis liquid chromatography tandem mass spectrometry (LC-MS/MS) method revealed lower levels of FT4 and FT3. Subsequent testing included heterophile blocking tube treatment, polyethylene glycol (PEG) precipitation, and anti-T3/T4 autoantibody levels. The tests revealed thyroid hormone autoantibodies (THAAs) as the cause of immunoassay interference.

CONCLUSIONS

When thyroid hormones are elevated and TSH is not suppressed, confirmatory testing with another method such as equilibrium dialysis LC-MS/MS, which is not susceptible to interference from autoantibodies, should be considered.

Для цитирования: Goettemoeller Travis, McShane Adam J, Rao Pratibha. Misleading FT4 and FT3 Due to Immunoassay Interference From Autoantibodies.. Clinical biochemistry. 2021. DOI: 10.1016/j.clinbiochem.2021.11.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.