High-throughput quantification of lysophosphatidylcholine by electrospray ionization tandem mass spectrometry.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Liebisch Gerhard, Drobnik Wolfgang, Lieser Bernd, Schmitz Gerd
Аннотация
BACKGROUND: Lysophosphatidylcholine (LPC) has been suggested to play a functional role in various diseases, including atherosclerosis, diabetes, and cancer mediated by LPC-specific G-protein-coupled receptors. Initial studies provided evidence for a potential use of LPC as diagnostic maker. However, existing methodologies are of limited value for a systematic evaluation of LPC species concentrations because of complicated, time-consuming procedures. We describe a methodology based on electrospray ionization tandem mass spectrometry (ESI-MS/MS) applicable for high-throughput LPC quantification.
METHODS: Crude lipid extracts of EDTA-plasma samples were used for direct flow injection analysis. LPC 13:0 and LPC 19:0 were added as internal standards, and the ESI-MS/MS was operated in the parent-scan mode for m/z 184. Quantification was achieved by standard addition. Data processing was highly automated by use of the mass spectrometer software and self-programmed Excel macros.
RESULTS: The calibrators LPC 16:0, LPC 18:0, and LPC 22:0 showed a linear response independent of sample dilution and plasma cholesterol concentration for both internal standards. The within-run imprecision (CV) was 3% for the major and 12% for the minor species, whereas the total imprecision was approximately 12% for the major and 25% for the minor species. The detection limit was <1 micromol/L.
CONCLUSION: The developed ESI-MS/MS methodology with an analysis time of 2 min/sample, simple sample preparation, and automated data analysis allows high-throughput quantification of distinct LPC species from plasma samples, which could be a valuable tool for the evaluation of LPC as diagnostic marker.
METHODS: Crude lipid extracts of EDTA-plasma samples were used for direct flow injection analysis. LPC 13:0 and LPC 19:0 were added as internal standards, and the ESI-MS/MS was operated in the parent-scan mode for m/z 184. Quantification was achieved by standard addition. Data processing was highly automated by use of the mass spectrometer software and self-programmed Excel macros.
RESULTS: The calibrators LPC 16:0, LPC 18:0, and LPC 22:0 showed a linear response independent of sample dilution and plasma cholesterol concentration for both internal standards. The within-run imprecision (CV) was 3% for the major and 12% for the minor species, whereas the total imprecision was approximately 12% for the major and 25% for the minor species. The detection limit was <1 micromol/L.
CONCLUSION: The developed ESI-MS/MS methodology with an analysis time of 2 min/sample, simple sample preparation, and automated data analysis allows high-throughput quantification of distinct LPC species from plasma samples, which could be a valuable tool for the evaluation of LPC as diagnostic marker.