DOI 10.1007/s12350-019-01621-2

Медицинская библиотека · Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2019

Myocardial blood flow quantitation in patients with congestive heart failure: head-to-head comparison between rapid-rotating gantry SPECT and CZT SPECT.

Ma Rongzheng, Wang Lei, Wu Dayong, Wang Meng, Sun Xiaoxin, Hsu Bailing, Fang Wei

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Recently, the feasibility of myocardial blood flow (MBF) quantitation using rapid-rotating gantry (RRG) and cadmium-zinc-telluride (CZT) SPECT cameras has been demonstrated. We compared MBF quantitation using these two camera systems.

METHODS

Twenty patients with congestive heart failure (CHF) and 20 patients without CHF (non-CHF) were included. On two consecutive days, dynamic SPECT imaging was performed after a bolus injection of 20 mCi of 99mTc-Sestamibi (MIBI) with RRG-SPECT and list-mode acquisition with CZT-SPECT. All dynamic SPECT images were reconstructed with full physical corrections, corrections for ventricular spillover and partial volume effect, using one-tissue kinetic modeling. Resting MBF converted from K1 was then corrected for MIBI extraction fraction and adjusted for rate-pressure product.

RESULTS

In both patient groups, there was no significant difference between resting MBF values measured with RRG-SPECT and CZT-SPECT systems (P = 0.06, P = 0.2 respectively). For CHF patients, linear regression (LR) was y(RRG) = 1.0412x (CZT) (r = 0.97) with a small systemic difference (Δ = 0.03 mL·min-1·g-1, 95% CI - 0.11 to 0.20) by Bland-Altman analysis. For non-CHF patients, LR was y(RRG) = 1.025x (CZT) (r = 0.89) with also small systemic difference (ΔT= 0.02 mL·min-1·g-1, 95% CI - 0.14 to 0.19) in BA analysis.

CONCLUSION

Physical corrections along with other image corrections can provide comparable MBF quantitations in both CHF and non-CHF patients, regardless of the type of SPECT systems used.

Для цитирования: Ma Rongzheng, Wang Lei, Wu Dayong, Wang Meng, Sun Xiaoxin, Hsu Bailing, Fang Wei. Myocardial blood flow quantitation in patients with congestive heart failure: head-to-head comparison between rapid-rotating gantry SPECT and CZT SPECT.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2019. DOI: 10.1007/s12350-019-01621-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.