Latent Pulmonary Vascular Disease May Alter the Response to Therapeutic Atrial Shunt Device in Heart Failure.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Borlaug Barry A, Blair John, Bergmann Martin W, Bugger Heiko, Burkhoff Dan, Bruch Leonhard, Celermajer David S, Claggett Brian, Cleland John G F, Cutlip Donald E, Dauber Ira, Eicher Jean-Christophe, Gao Qi, Gorter Thomas M, Gustafsson Finn, Hayward Chris, van der Heyden Jan, Hasenfuß Gerd, Hummel Scott L, Kaye David M, Komtebedde Jan, Massaro Joseph M, Mazurek Jeremy A, McKenzie Scott, Mehta Shamir R, Petrie Mark C, Post Marco C, Nair Ajith, Rieth Andreas, Silvestry Frank E, Solomon Scott D, Trochu Jean-Noël, Van Veldhuisen Dirk J, Westenfeld Ralf, Leon Martin B, Shah Sanjiv J, REDUCE LAP-HF-II Investigators
Аннотация
METHODS: REDUCE LAP-HF II enrolled 626 patients with heart failure, ejection fraction ≥40%, exercise pulmonary capillary wedge pressure ≥25 mm Hg, and resting pulmonary vascular resistance <3.5 Wood units who were randomized 1:1 to atrial shunt device or sham control. The primary outcome-a hierarchical composite of cardiovascular death, nonfatal ischemic stroke, recurrent HF events, and change in health status-was analyzed using the win ratio. Latent PVD was defined as pulmonary vascular resistance ≥1.74 Wood units (highest tertile) at peak exercise, measured before randomization.
RESULTS: Compared with patients without PVD (n=382), those with latent PVD (n=188) were older, had more atrial fibrillation and right heart dysfunction, and were more likely to have elevated left atrial pressure at rest. Shunt treatment was associated with worse outcomes in patients with PVD (win ratio, 0.60 [95% CI, 0.42, 0.86]; P=0.005) and signal of clinical benefit in patients without PVD (win ratio, 1.31 [95% CI, 1.02, 1.68]; P=0.038). Patients with larger right atrial volumes and men had worse outcomes with the device and both groups were more likely to have pacemakers, heart failure with mildly reduced ejection fraction, and increased left atrial volume. For patients without latent PVD or pacemaker (n=313; 50% of randomized patients), shunt treatment resulted in more robust signal of clinical benefit (win ratio, 1.51 [95% CI, 1.14, 2.00]; P=0.004).
CONCLUSIONS: In patients with heart failure with preserved or mildly reduced ejection fraction, the presence of latent PVD uncovered by invasive hemodynamic exercise testing identifies patients who may worsen with atrial shunt therapy, whereas those without latent PVD may benefit.