Real-time CMR guidance for intracardiac and great vessel pressure mapping in patients with congenital heart disease using an MR conditional guidewire-results of 25 patients.

Meierhofer Christian, Belker Kristina, Shehu Nerejda, Latus Heiner, Mkrtchyan Naira, Naumann Susanne, Martinoff Stefan, Stern Heiko, Eicken Andreas, Ewert Peter

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The aim of this study was to test a CE-certified MR-conditional guidewire to facilitate blood pressure measurement in cardiovascular magnetic resonance (CMR) using fluid-filled catheters in patients with congenital heart disease (CHD). The main purpose was to determine procedural success in a post market clinical follow-up (PMCF) for routine procedure in a diagnostic and interventional workflow. Real-time CMR provides high quality imaging without the risk of exposing the patient to X-rays, especially for patients with irregular heart anatomy and patients who are susceptible to radiation and iodinated contrast media. To date, the assessment of blood pressure gradients is not a common feature of CMR, as these gradients cannot be accurately evaluated in routine CMR.

METHODS

Twenty-five CHD patients who were planned for combined clinical CMR and diagnostic and/or interventional catheterization were enrolled in the trial. Prior to inclusion, a specific procedure for catheterization in CMR was defined, encompassing the assessment of pressure and pressure gradients in the heart and great vessels.

RESULTS

By the use of an MR-conditional guidewire we successfully measured specific pressure and pressure gradients in up to 92% of cases with liquid-filled catheters which were guided exclusively under CMR guidance. There were no guidewire-related adverse events, and guidewire guidance and manipulation of catheters were successful.

CONCLUSIONS

Using a MR-conditional guidewire assists in easily reaching targets in the heart and great vessels and makes the catheter itself visible, so that invasive blood pressure assessment by CMR guidance with liquid-filled catheters can be improved.

KEYWORDS: Cardiovascular magnetic resonance (CMR); congenital heart disease (CHD); cardiac catheterization; magnetic resonance; pressure; guidewire.

Для цитирования: Meierhofer Christian, Belker Kristina, Shehu Nerejda, Latus Heiner, Mkrtchyan Naira, Naumann Susanne, Martinoff Stefan, Stern Heiko, Eicken Andreas, Ewert Peter. Real-time CMR guidance for intracardiac and great vessel pressure mapping in patients with congenital heart disease using an MR conditional guidewire-results of 25 patients.. Cardiovascular diagnosis and therapy. 2021. DOI: 10.21037/cdt-20-575.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.