OBJECTIVES: Cardiac output (CO) measurements employing the direct Fick principle represent the gold standard in right-sided heart catheterization (RHC). The current widespread approach in hemodynamic workup however uses the indirect Fick principle with assumed values for oxygen uptake (VO2) leading to incorrect CO values in up to 25% of patients. We have tested a contemporary breath-by-breath gas analyzer that allows precise real-time measurements of VO2 with appropriate time and effort to serve the direct Fick principle.
METHODS: By means of a small and mobile metabolic cart assembled with widely used components of a standard spiroergometer, we performed bedside measurements of individual VO2. In 33 unselected, consecutive patients with various indications for RHC we compared CO values derived from indirect vs. direct Fick calculations.
RESULTS: In 28 of the 33 patients, VO2 measurements were completed with a plausible dataset within a median of 3.2 (interquartile range 2.8-6.2) min. In nine of the 28 patients, CO values based on measured VO2 values differed by more than 20% from CO calculations based on assumed VO2 values with value deviations scattering over a broad range in both directions (maximally +52% to minimally -46%).
CONCLUSIONS: The bedside measurement of VO2 for gold standard CO determination is technically feasible within a few min and can thus be easily included in any RHC protocol. As modern therapy for numerus indications demand a precise upfront measurement of hemodynamics, our method might help to correctly identify patients for costly therapies.
METHODS: By means of a small and mobile metabolic cart assembled with widely used components of a standard spiroergometer, we performed bedside measurements of individual VO2. In 33 unselected, consecutive patients with various indications for RHC we compared CO values derived from indirect vs. direct Fick calculations.
RESULTS: In 28 of the 33 patients, VO2 measurements were completed with a plausible dataset within a median of 3.2 (interquartile range 2.8-6.2) min. In nine of the 28 patients, CO values based on measured VO2 values differed by more than 20% from CO calculations based on assumed VO2 values with value deviations scattering over a broad range in both directions (maximally +52% to minimally -46%).
CONCLUSIONS: The bedside measurement of VO2 for gold standard CO determination is technically feasible within a few min and can thus be easily included in any RHC protocol. As modern therapy for numerus indications demand a precise upfront measurement of hemodynamics, our method might help to correctly identify patients for costly therapies.