The British journal of radiology · 2022

The use of digital magnification to reduce radiation dose in the cardiac catheter laboratory.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Dalvi Shailesh, Roberts Hywel Mortimer, Bellamy Christopher, Rees Michael

Аннотация

OBJECTIVES: To audit whether using magnification of images by use of a large viewing screen using digital matrix magnification which enlarges the image by 33% without using the X-ray machine zoom magnification protocols on a Siemens Artis Zee X-ray machine in a cardiac catheter laboratory results in a reduction of kerma-area product (KAP) for both diagnostic and interventional procedures. This reduction was predicted in an in vitro study in our laboratory, which has previously shown a 20.4% reduction in KAP.

METHODS: A retrospective analysis was conducted of the radiation exposure to compare the measured KAP recorded during the period when conventional magnification with automatic brightness and dose control was used on a Siemens Artis Zee X-ray machine with a flat panel detector and when magnification settings were avoided by using a large screen to enlarge and project a non-magnified image by digital magnification. The analysis was carried out for patients having a diagnostic coronary angiogram and those having an interventional coronary procedure.

RESULTS: For diagnostic coronary angiograms the median KAP per procedure in the period using conventional magnification was 2124.5 µGy.m2 compared to 1401 µGy.m2 when image matrix magnification was used, a 34% reduction (p < 0.0001). For interventional coronary procedures, the median KAP per procedure in the period using conventional magnification was 3791 µGy.m2 compared to 2568.5 µGy.m2 when image matrix magnification was used, a 32% reduction (p < 0.0001).

CONCLUSION: Avoiding using conventional magnification in the cardiac catheter laboratory and using a large screen to magnify images was associated with a statistically significant greater than 30% reduction in KAP.

ADVANCES IN KNOWLEDGE: This paper is the proof in clinical practice of a theoretical conclusion that radiation dose (KAP) is reduced by use of Image matrix magnification using a large viewing screen without the need to use X-ray tube magnification without significant loss of image resolution in interventional cardiology. The same approach will be useful in interventional radiology.