DOI 10.1259/bjr.20210269

Медицинская библиотека · The British journal of radiology · 2022

The use of digital magnification to reduce radiation dose in the cardiac catheter laboratory.

Dalvi Shailesh, Roberts Hywel Mortimer, Bellamy Christopher, Rees Michael

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To audit whether using magnification of images by use of a large viewing screen using digital matrix magnification which enlarges the image by 33% without using the X-ray machine zoom magnification protocols on a Siemens Artis Zee X-ray machine in a cardiac catheter laboratory results in a reduction of kerma-area product (KAP) for both diagnostic and interventional procedures. This reduction was predicted in an in vitro study in our laboratory, which has previously shown a 20.4% reduction in KAP.

METHODS

A retrospective analysis was conducted of the radiation exposure to compare the measured KAP recorded during the period when conventional magnification with automatic brightness and dose control was used on a Siemens Artis Zee X-ray machine with a flat panel detector and when magnification settings were avoided by using a large screen to enlarge and project a non-magnified image by digital magnification. The analysis was carried out for patients having a diagnostic coronary angiogram and those having an interventional coronary procedure.

RESULTS

For diagnostic coronary angiograms the median KAP per procedure in the period using conventional magnification was 2124.5 µGy.m2 compared to 1401 µGy.m2 when image matrix magnification was used, a 34% reduction (p < 0.0001). For interventional coronary procedures, the median KAP per procedure in the period using conventional magnification was 3791 µGy.m2 compared to 2568.5 µGy.m2 when image matrix magnification was used, a 32% reduction (p < 0.0001).

CONCLUSION

Avoiding using conventional magnification in the cardiac catheter laboratory and using a large screen to magnify images was associated with a statistically significant greater than 30% reduction in KAP.

ADVANCES IN KNOWLEDGE: This paper is the proof in clinical practice of a theoretical conclusion that radiation dose (KAP) is reduced by use of Image matrix magnification using a large viewing screen without the need to use X-ray tube magnification without significant loss of image resolution in interventional cardiology. The same approach will be useful in interventional radiology.

Для цитирования: Dalvi Shailesh, Roberts Hywel Mortimer, Bellamy Christopher, Rees Michael. The use of digital magnification to reduce radiation dose in the cardiac catheter laboratory.. The British journal of radiology. 2022. DOI: 10.1259/bjr.20210269.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.