m+Medelevator.RU
← Поиск публикаций
Veterinary anaesthesia and analgesia · 2024

Evaluation of a novel, low-cost, 3D printed video laryngoscope with borescope in anesthetized Beagle dogs.

СтатьяОценочное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Son Won-Gyun, Sung Taehoon, Shin Donghwi, Rhee Suehyung, Nam Changhoon, Kim Minha, Park Chailin, Lee Jungha, Kim Junsoo, Lee Inhyung

Аннотация

OBJECTIVE: To develop and evaluate a low-cost three-dimensional (3D)-printed video laryngoscope (VLVET) for use with a commercial borescope.

STUDY DESIGN: Instrument development and pilot study.

ANIMALS: A total of six adult male Beagle dogs.

METHODS: The VLVET consisted of a laryngoscope handle and a Miller-type blade, and a detachable camera holder that attached to various locations along the blade. The laryngoscope and camera holder were 3D-printed using black polylactic acid filament. Dogs were premedicated with intravenous (IV) medetomidine (15 μg kg-1) and anesthesia induced with IV alfaxalone (1.5 mg kg-1). The VLVET, combined with a borescope, was used for laryngeal visualization and intubation. Performance was evaluated by comparing direct and video-assisted views in sternal recumbency. The borescope camera was sequentially positioned at 2, 4, 6, 8 and 10 cm from the blade tip (distanceLARYNX-CAM), which was placed on the epiglottis during intubation or laryngoscopy. At the 10 cm distanceLARYNX-CAM, laryngeal visualization was sequentially scored at inter-incisor gaps of 10, 8, 6, 4 and 2 cm. Laryngeal visualization scores (0-3 range, with 0 = obstructed and 3 = unobstructed views) were statistically analyzed using the Friedman's test.

RESULTS: Under direct visualization, the 2 cm distanceLARYNX-CAM had a significantly lower score compared with all other distanceLARYNX-CAM (all p = 0.014) because the view was obstructed by the camera holder and borescope camera. With both direct and camera-assisted views, visualization scores were higher at inter-incisor gaps ≥ 4 cm compared with 2 cm (all p < 0.05).

CONCLUSIONS AND CLINICAL RELEVANCE: During laryngoscopy and intubation, the VLVET and borescope facilitated both direct and video laryngoscopy at distanceLARYNX-CAM in Beagle dogs when inter-incisor gaps were ≥ 4 cm.