Ultrasound assessment of bladder volume: is it valid after delivery?

Pallis Lucinda M, Wilson Melissa

Сравнительное исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

AIM: To assess the validity of a commercially available bladder volume scanner in the puerperium.

METHODS

A prospective blinded comparison of bladder volume measurement after vaginal delivery using the BladderScan bladder volume instrument (BVI) 3000 and Foley catheter; comparison using the intraclass correlation coefficient.

RESULTS

The mean difference between the two measurements was a 130-mL over-measurement by the bladder scanner (range: -156 mL to +422 mL). The intraclass correlation coefficient was 0.23 (95% confidence interval 0.00, 0.59).

CONCLUSIONS

The BladderScan BVI 3000 is not an accurate instrument to assess bladder volume the day after vaginal delivery. Some of the discrepancy might relate to use of the Foley catheter as the reference standard. Further comparison between the BladderScan and a short female catheter or real time ultrasound is indicated.

Для цитирования: Pallis Lucinda M, Wilson Melissa. Ultrasound assessment of bladder volume: is it valid after delivery?. The Australian & New Zealand journal of obstetrics & gynaecology. 2003. DOI: 10.1046/j.0004-8666.2003.00141.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.