Reference ranges of late-night salivary cortisol and cortisone measured by LC-MS/MS and accuracy for the diagnosis of Cushing's syndrome.
Аннотация
PURPOSE: International guidelines recommend salivary cortisol for the diagnosis of Cushing's syndrome. Despite mass spectrometry-based assays are considered the analytical gold-standard, there is still the need to define reference intervals and diagnostic accuracy of such methodology.
METHODS
100 healthy volunteers and 50 consecutive patients were enrolled to compare LC-MS/MS and electrochemiluminescence assay for the determination of late-night salivary cortisol and cortisone. Moreover, we aimed to determine reference intervals of salivary steroids in a population of healthy individuals and diagnostic accuracy in patients with suspected hypercortisolism and in a population including also healthy individuals.
RESULTS
Method comparison highlighted a positive bias (51.8%) of immunoassay over LC-MS/MS. Reference intervals of salivary cortisol (0.17-0.97 µg/L), cortisone (0.84-4.85 µg/L) and ratio (0.08-0.30) were obtained. The most accurate thresholds of salivary cortisol for the diagnosis of hypercortisolism were 1.15 µg/L in the population with suspected hypercortisolism (AUC 1) and 1.30 µg/L in the population including also healthy individuals (AUC 1). Cut-off values of salivary cortisone (7.23 µg/L; Se 92.9%, Sp 97.2%, AUC 0.960 and Se 92.9%, Sp 99.1%, AUC 0.985 in suspected hypercortisolism and in overall population, respectively) and cortisol-to-cortisone ratio (0.20; Se 85.7%, Sp 80.6%, AUC 0.820 and Se 85.7%, Sp 85.5%, AUC 0.855 in suspected hypercortisolism and in overall population, respectively) were accurate and similar in both populations.
CONCLUSION
LC-MS/MS is the most accurate analytical platform for measuring salivary steroids. Obtained reference intervals are coherent with previously published data and diagnostic accuracy for diagnosis of overt hypercortisolism proved highly satisfactory.