DOI 10.1007/s10877-017-9979-0

Медицинская библиотека · Journal of clinical monitoring and computing · 2017

Influence of the marvelous™ three-way stopcock on the natural frequency and damping coefficient in blood pressure transducer kits.

Fujiwara Shigeki Joseph Luke, Tachihara Keiichi, Mori Satoshi, Ouchi Kentaro, Itakura Shoko, Yasuda Michiko, Hitosugi Takashi, Imaizumi Uno, Miki Yoichiro, Toyoguchi Izumi, Yoshida Kazu-Ichi, Yokoyama Takeshi

СтатьяПеревод пока не подготовлен

Аннотация

Two types of Planecta™ ports are commonly used as sampling ports in blood pressure transducer kits: a flat-type port (FTP) and a port with a three-way stopcock (PTS). Recently, a new type of three-way stopcock (Marvelous™) has been released as a Planecta™ counterpart, but its effects on the frequency characteristics and reliability of blood pressure monitoring have not been investigated. We assessed the influence of the Marvelous™ stopcock on the frequency characteristics of the pressure transducer kit. The basic pressure transducer kit, DT4812J, was modified by replacing one or two of the original three-way stopcocks with Marvelous™ stopcocks. The frequency characteristics (i.e., natural frequency and damping coefficient) of each kit were determined using wave parameter analysis software, and subsequently evaluated on a Gardner chart. Replacement of the original blood pressure transducer kit stopcocks with Marvelous™ stopcocks decreased the natural frequency (48.3 Hz) to 46.3 Hz or 44.8 Hz, respectively; the damping coefficient was not significantly changed. Plotting the data on a Gardner chart revealed that the changes fell within the adequate dynamic response region, indicating they were within the allowable range. Insertion of Marvelous™ stopcocks slightly affects the natural frequency of the pressure transducer kit, similar to inserting a PTS. The results indicate that the Marvelous™ stopcock is useful for accurate monitoring of arterial blood pressure, and may be recommended when insertion of two or more closed-loop blood sampling systems is necessary.

Для цитирования: Fujiwara Shigeki Joseph Luke, Tachihara Keiichi, Mori Satoshi, Ouchi Kentaro, Itakura Shoko, Yasuda Michiko, Hitosugi Takashi, Imaizumi Uno, Miki Yoichiro, Toyoguchi Izumi, Yoshida Kazu-Ichi, Yokoyama Takeshi. Influence of the marvelous™ three-way stopcock on the natural frequency and damping coefficient in blood pressure transducer kits.. Journal of clinical monitoring and computing. 2017. DOI: 10.1007/s10877-017-9979-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.