Reuse of internal pulse generator in cases of infection after deep brain stimulation surgery.
Аннотация
BACKGROUND
Hardware-related infection after deep brain stimulation (DBS) is one of the most serious complications and may need additional interventions.
OBJECTIVES
To reuse the internal pulse generator (IPG) after DBS infection and to reduce the economic costs.
METHODS
A database of 102 patients who underwent DBS surgery was used in the study. The incidence, clinical characteristics and management of infections while reusing the IPG after DBS-related infection were analyzed and reported.
RESULTS
The overall infection rate was 5.9% (6 of 102 patients). Management consisted of total hardware removal followed by intravenous antibiotics. The IPG was at first kept in a solution, then rinsed with water and dried following sterilization with ethylene oxide gas at 38 °C for 18 h. When the treatment of the infection was finished, we reused the IPG and reimplanted the DBS. No hardware-related infection or other complications were observed after reimplantation.
CONCLUSIONS
Management of hardware-related infections can be challenging. The medical and economic costs associated with these infections are enormous. The IPG can often be saved in infected patients. Thus, a significant cost burden is eliminated. Properly executed, reuse of IPG should markedly reduce the costs of these devices.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.