Аннотация

AIMS: To evaluate passive-fixation lead failure rates and long-term patient survival in subjects implanted with Sprint Fidelis electrodes.

METHODS AND

RESULTS

We identified 748 subjects who received a Sprint Fidelis (n = 429; Medtronic models 6948: 94.8%, 6949: 2.6%, 6930: 1.9%, 6931: 0.7%) or a Sprint 'non-Fidelis' implantable cardioverter defibrillator lead (n = 319, Medtronic models 6944: 68.6%, 6947: 17.9%, 6942: 7.8%, 6943: 3.4%, 6945: 2.2%) at our centre between 1998 and 2008. Kaplan-Meier patient survival was lower in the Fidelis group than in the Control cohort (68.4 vs. 77.0% at 5 years, P = 0.0061), but multivariate analyses revealed no significant association between mortality and implanted lead type. Passive-fixation lead failure rate at 5 years was 14.4% (95% confidence interval (CI) [9.2, 19.3]) in the Fidelis (n = 414) group and 1.8% (95% CI [0.0-3.8]) in the Control (n = 241) cohort (P < 0.001 upon multivariate comparison).

CONCLUSION

Failure rates of passive-fixation Sprint Fidelis leads are increased and similar to those previously reported for active-fixation Fidelis electrodes. Despite the elevated risk for lead failure and its potential sequelae, the Sprint Fidelis has no obvious impact on long-term mortality.

Для цитирования: Vollmann Dirk, Woronowicz Sandra, Kmiec Lukasz, Jung Klaus, Zenker Dieter, Seegers Joachim, Sossalla Samuel, Dorenkamp Marc, Sohns Christian, Lüthje Lars, Hasenfuss Gerd, Zabel Markus. Passive-fixation lead failure rates and long-term patient mortality in subjects implanted with Sprint Fidelis electrodes.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2013. DOI: 10.1093/europace/eut185.