BACKGROUND

There has been increasing concern of metal-on-metal (MOM) hip replacements regarding adverse reactions to metal debris. Information regarding prevalence and risk factors for these adverse reactions is scarce.

QUESTIONS/PURPOSES: The primary purposes of our study were to determine (1) the prevalence of adverse reactions to metal debris among patients who received small-headed (< 50 mm) Articular Surface Replacement (ASR™) prostheses in hip resurfacing procedures or the ASR™ XL prostheses during THAs at our institution, and (2) the risk factors for adverse reactions to metal debris and if they are different in hip resurfacing replacements compared with THAs?

METHODS

Small-headed ASR™ prostheses were used in 482 operations (424 patients) at our institution. After the recall of ASR™ prostheses, we established a systematic screening program to find patients with adverse reactions to metal debris. At a mean of 4.9 years (range, 0.2-8.1 years) postoperatively, 379 patients (435 hips) attended a screening program, which consisted of clinical evaluation, whole blood cobalt and chromium measurements, and cross-sectional imaging.

RESULTS

At followup, 162 hips (34%) have been revised. The majority (85%) were revised owing to causes related to adverse reactions to metal debris. The 7-year survivorship was 51% for the ASR™ hip replacement cohort and 38% for the ASR™ XL THA cohort, respectively. Reduced cup coverage was an independent risk factor for adverse reactions to metal debris in both cohorts. High preoperative ROM, use of the Corail(®) stem, and female gender were associated with an increased risk of adverse reactions to metal debris only in patients undergoing THA.

CONCLUSIONS

Adverse reactions to metal debris are common with small-headed ASR™ prostheses. Risk factors for these adverse reactions differ between hip resurfacing procedures and THAs. Our results suggest a more complicated failure mechanism in THAs than in hip resurfacing procedures.

Для цитирования: Reito Aleksi, Puolakka Timo, Elo Petra, Pajamäki Jorma, Eskelinen Antti. High prevalence of adverse reactions to metal debris in small-headed ASR™ hips.. Clinical orthopaedics and related research. 2013. DOI: 10.1007/s11999-013-3023-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.