m+Medelevator.RU
← Поиск публикаций
Advances in radiation oncology · 2025

Impact of Injection Technique on Microsphere Distribution During Transarterial Radioembolization in a Successively Bifurcating In Vitro Model.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Snoeijink Tess J, van der Hoek Jan L, Brekel Anne van den, van Wolfswinkel Gerhard, Janssen Marcel J R, Jebbink Erik Groot, Nijsen J Frank W

Аннотация

PURPOSE: This study aimed to experimentally investigate how injection technique affects the distribution of microspheres during transarterial radioembolization in a successively bifurcating in vitro model.

METHODS AND MATERIALS: A symmetrical phantom, bifurcating 3 times into 8 outlets, was incorporated into a flow circuit. A blood-mimicking fluid was pumped through the phantom using a physiological representative waveform. A microcatheter was placed into the lumen of the phantom, and holmium-165 microspheres were administered with a conventional administration device and a newly designed controlled administration device, containing a rotating syringe to keep the microspheres in suspension during administration. Two clinicians performed manual injections to establish clinically relevant injection rates. Then, different injection profiles were tested using syringe pumps: pulsed vs continuous injections (24 mL/min), and reduced continuous injection rates (10 and 5 mL/min). Microspheres were collected at each outlet and their distribution over the 8 outlets was analyzed.

RESULTS: Continuous high injection rates led to more homogeneous radial distributions of microspheres over the right side of the phantom (outlet 5-8 received 16.5%-23.1% of the microspheres per outlet) compared with the clinically standard used pulsed injections (outlet 5-8 received 11.3%-40.1% of the microspheres per outlet). In contrast, reduced continuous injection rates led to more selective distributions (outlet 5-8 received 2.5%-68.8% of the microspheres at 10 mL/min and 1.0%-80.0% at 5 mL/min).

CONCLUSIONS: Injection technique strongly influences the distribution of microspheres. During high continuous injections, more mixing between microspheres and blood-mimicking fluid was observed. This led to more uniform radial microsphere distributions, creating a more predictive setting for transarterial radioembolization.

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты