DOI 10.1136/bmjinnov-2021-000845

Медицинская библиотека · BMJ innovations · 2022

Design and proof-of-concept evaluation of a touchless connector system for preventing peritoneal dialysis-associated peritonitis.

Yekinni Ibrahim O, Viker Tom, Hunter Ryan, Tucker Aaron, Elfering Sarah, Rheault Michelle N, Erdman Arthur

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

In this paper, we describe the design of a touchless peritoneal dialysis connector system and how we evaluated its potential for preventing peritoneal dialysis-associated peritonitis, in comparison to the standard of care. The unique feature of this system is an enclosure within which patients can connect and disconnect for therapy, protecting their peritoneal catheters from touch or aerosols.

METHODS

We simulated a worst-case contamination scenario by spraying 40mL of a standardized inoculum [ 1×107 colony-forming units (CFU) per milliliter] of test organisms, Staphylococcus epidermidis ATCC1228 and Pseudomonas aeruginosa ATCC39327, while test participants made mock connections for therapy. We then compared the incidence of fluid path contamination by test organisms in the touchless connector system and the standard of care. 4 participants were recruited to perform a total of 56 tests, divided in a 1:1 ratio between both systems. Peritoneal dialysis fluid sample from each test was collected and maintained at body temperature (37° C) for 16 hours before being plated on Luria Bertani agar, Mannitol Salts Agar and Pseudomonas isolation agar for enumeration.

RESULTS

No contamination was observed in test samples from the touchless connector system, compared to 65%, 75% and 70% incidence contamination for the standard of care on Luria Bertani agar, Mannitol Salts Agar and Pseudomonas isolation agar respectively.

CONCLUSION

Results show that the touchless connector system can prevent fluid path contamination even in heavy bacterial exposures and may help reduce peritoneal dialysis-associated peritonitis risks from inadvertent contamination with further development.

Для цитирования: Yekinni Ibrahim O, Viker Tom, Hunter Ryan, Tucker Aaron, Elfering Sarah, Rheault Michelle N, Erdman Arthur. Design and proof-of-concept evaluation of a touchless connector system for preventing peritoneal dialysis-associated peritonitis.. BMJ innovations. 2022. DOI: 10.1136/bmjinnov-2021-000845.