Evaluation of a New Extracorporeal CO2 Removal Device in an Experimental Setting.

Di Nardo Matteo, Annoni Filippo, Su Fuhong, Belliato Mirko, Lorusso Roberto, Broman Lars Mikael, Malfertheiner Maximilian, Creteur Jacques, Taccone Fabio Silvio

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Ultra-protective lung ventilation in acute respiratory distress syndrome or early weaning and/or avoidance of mechanical ventilation in decompensated chronic obstructive pulmonary disease may be facilitated by the use of extracorporeal CO2 removal (ECCO2R). We tested the CO2 removal performance of a new ECCO2R (CO2RESET) device in an experimental animal model.

METHODS

Three healthy pigs were mechanically ventilated and connected to the CO2RESET device (surface area = 1.8 m2, EUROSETS S.r.l., Medolla, Italy). Respiratory settings were adjusted to induce respiratory acidosis with the adjunct of an external source of pure CO2 (target pre membrane lung venous PCO2 (PpreCO2): 80-120 mmHg). The amount of CO2 removed (VCO2, mL/min) by the membrane lung was assessed directly by the ECCO2R device.

RESULTS

Before the initiation of ECCO2R, the median PpreCO2 was 102.50 (95.30-118.20) mmHg. Using fixed incremental steps of the sweep gas flow and maintaining a fixed blood flow of 600 mL/min, VCO2 progressively increased from 0 mL/min (gas flow of 0 mL/min) to 170.00 (160.00-200.00) mL/min at a gas flow of 10 L/min. In particular, a high increase of VCO2 was observed increasing the gas flow from 0 to 2 L/min, then, VCO2 tended to progressively achieve a steady-state for higher gas flows. No animal or pump complications were observed.

CONCLUSIONS

Medium-flow ECCO2R devices with a blood flow of 600 mL/min and a high surface membrane lung (1.8 m2) provided a high VCO2 using moderate sweep gas flows (i.e., >2 L/min) in an experimental swine models with healthy lungs.

Для цитирования: Di Nardo Matteo, Annoni Filippo, Su Fuhong, Belliato Mirko, Lorusso Roberto, Broman Lars Mikael, Malfertheiner Maximilian, Creteur Jacques, Taccone Fabio Silvio. Evaluation of a New Extracorporeal CO2 Removal Device in an Experimental Setting.. Membranes. 2020. DOI: 10.3390/membranes11010008.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.