Медицинская библиотека · Anadolu kardiyoloji dergisi : AKD = the Anatolian journal of cardiology · 2007

Appraisal of BSPM obtained from the limited lead system.

Sobieszczanska Malgorzata, Jagielski Jozef, Nowak Barbara, Pilecki Witold, Kalka Dariusz

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Study goal was a comparative analysis of the characteristics of the isopotential maps, registered originally in the body surface potential mapping (BSPM) 87-lead (complete) cylindrical system, which were then transformed to the 30-lead (limited) spherical subsystem.

METHODS

The comparative studies were carried out on the electrocardiogram (ECG) recordings recorded originally in the 87-lead BSPM Fukuda Denshi system (HPM-7100; Fukuda Denshi, Co., Tokyo, Japan) in a group of 21 patients with right bundle branch blocks (RBBB) with the mean age of 61.3+/-10.2 years.

RESULTS

It was found that the body surface maps from the two multielectrode recording systems did not show any significant difference as to heart potential migration and distribution. It is worth noting that the maps created in the limited lead, spherical system gained the more realistic image, resulting from the effect of cardioelectric space symmetrization.

CONCLUSION

The results led to conclusion that a procedure of spherical transformation to the system with the reduced number of the recording electrodes did not influence the core features of the body surface heart potential maps.

Для цитирования: Sobieszczanska Malgorzata, Jagielski Jozef, Nowak Barbara, Pilecki Witold, Kalka Dariusz. Appraisal of BSPM obtained from the limited lead system.. Anadolu kardiyoloji dergisi : AKD = the Anatolian journal of cardiology. 2007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.